Laboratory Medicine 2017 August 2
Background: International normalized ratio (INR) derivation is dependent on the international sensitivity index (ISI) of thromboplastin. It varies with instrument and reagents used.

Objective: To evaluate the role of a correction factor in the derivation of INR.

Methods: We studied prothrombin time (PT) and INR from patients using 3 thromboplastins of varying ISI values. The correction factor was applied to the observed INR to obtain a corrected INR.

Results: The difference between corrected INR and observed INR values varies from -0.8 through 0.96.

Conclusions: Corrected INR is dependent on PT only. It can be applied to all patients irrespective of cause for elevated PT.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app