Sensitivity and specificity of two different automated external defibrillators.
Resuscitation 2017 November
AIM: The aim was to investigate the clinical performance of two different types of automated external defibrillators (AEDs).
METHODS: Three investigators reviewed 2938 rhythm analyses performed by AEDs in 240 consecutive patients (median age 72, q1-q3=62-83) who had suffered cardiac arrest between January 2011 and March 2015. Two different AEDs were used (AED A n=105, AED B n=135) in-hospital (n=91) and out-of-hospital (n=149).
RESULTS: Among 194 shockable rhythms, 17 (8.8%) were not recognized by AED A, while AED B recognized 100% (n=135) of shockable episodes (sensitivity 91.2 vs 100%, p<0.01). In AED A, 8 (47.1%) of these episodes were judged to be algorithm errors while 9 (52.9%) were caused by external artifacts. Among 1039 non-shockable rhythms, AED A recommended shock in 11 (1.0%), while AED B recommended shock in 63 (4.1%) of 1523 episodes (specificity 98.9 vs 95.9, p<0.001). In AED A, 2 (18.2%) of these episodes were judged to be algorithm errors (AED B, n=40, 63.5%), while 9 (81.8%) were caused by external artifacts (AED B, n=23, 36.5%).
CONCLUSIONS: There were significant differences in sensitivity and specificity between the two different AEDs. A higher sensitivity of AED B was associated with a lower specificity while a higher specificity of AED A was associated with a lower sensitivity. AED manufacturers should work to improve the algorithms. In addition, AED use should always be reviewed with a routine for giving feedback, and medical personnel should be aware of the specific strengths and shortcomings of the device they are using.
METHODS: Three investigators reviewed 2938 rhythm analyses performed by AEDs in 240 consecutive patients (median age 72, q1-q3=62-83) who had suffered cardiac arrest between January 2011 and March 2015. Two different AEDs were used (AED A n=105, AED B n=135) in-hospital (n=91) and out-of-hospital (n=149).
RESULTS: Among 194 shockable rhythms, 17 (8.8%) were not recognized by AED A, while AED B recognized 100% (n=135) of shockable episodes (sensitivity 91.2 vs 100%, p<0.01). In AED A, 8 (47.1%) of these episodes were judged to be algorithm errors while 9 (52.9%) were caused by external artifacts. Among 1039 non-shockable rhythms, AED A recommended shock in 11 (1.0%), while AED B recommended shock in 63 (4.1%) of 1523 episodes (specificity 98.9 vs 95.9, p<0.001). In AED A, 2 (18.2%) of these episodes were judged to be algorithm errors (AED B, n=40, 63.5%), while 9 (81.8%) were caused by external artifacts (AED B, n=23, 36.5%).
CONCLUSIONS: There were significant differences in sensitivity and specificity between the two different AEDs. A higher sensitivity of AED B was associated with a lower specificity while a higher specificity of AED A was associated with a lower sensitivity. AED manufacturers should work to improve the algorithms. In addition, AED use should always be reviewed with a routine for giving feedback, and medical personnel should be aware of the specific strengths and shortcomings of the device they are using.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app