Technical Note: A simple algorithm to convert EPID gray values into absorbed dose to water without prior knowledge.
Medical Physics 2017 December
PURPOSE: Full integration of EPID-based dosimetry in a global quality control workflow is still complicated. All the actual solutions are based on a relation between image gray-level signal and total linac-delivered dose. In this study, we propose a simple algorithm relying pixel gray-level of EPID image with average linac delivered dose per acquisition frame.
METHODS: Calibration models are constructed for Varian and Elekta linacs including scattering conditions and EPID-arm backscatter-specific corrections. Only simple homogeneous fields are required to establish the EPID dose conversion model for each x-ray beam. Then, the model was evaluated by comparing calculated and converted dose distributions for homogeneous and modulated beams using gamma maps.
RESULTS: To fit average dose per frame (Dfnorm ) vs pixel gray value (Ngnorm ) of each EPID image, a logarithmic curve Dfnorm=A+B∗lnNgnorm-C, has been chosen where A, B and C are constants depending on beam energy. Gamma comparison (2%, 2 mm, threshold 15%) between converted images and calculated dose distributions for linac control and pretreatment patient fields led to a gamma pass rate higher than 97% for all the analyzed fields.
CONCLUSIONS: Without a prior irradiation settings knowledge except the incident energy beam, we use EPID as a reliable dose to water detector for both homogeneous and modulated beams.
METHODS: Calibration models are constructed for Varian and Elekta linacs including scattering conditions and EPID-arm backscatter-specific corrections. Only simple homogeneous fields are required to establish the EPID dose conversion model for each x-ray beam. Then, the model was evaluated by comparing calculated and converted dose distributions for homogeneous and modulated beams using gamma maps.
RESULTS: To fit average dose per frame (Dfnorm ) vs pixel gray value (Ngnorm ) of each EPID image, a logarithmic curve Dfnorm=A+B∗lnNgnorm-C, has been chosen where A, B and C are constants depending on beam energy. Gamma comparison (2%, 2 mm, threshold 15%) between converted images and calculated dose distributions for linac control and pretreatment patient fields led to a gamma pass rate higher than 97% for all the analyzed fields.
CONCLUSIONS: Without a prior irradiation settings knowledge except the incident energy beam, we use EPID as a reliable dose to water detector for both homogeneous and modulated beams.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app