Cerebral Embolic Protection During Transcatheter Aortic Valve Replacement Significantly Reduces Death and Stroke Compared With Unprotected Procedures.

OBJECTIVES: The aim of this study was to evaluate the impact of cerebral embolic protection on stroke-free survival in patients undergoing transcatheter aortic valve replacement (TAVR).

BACKGROUND: Imaging data on cerebral embolic protection devices have demonstrated a significant reduction in number and volume of cerebral lesions.

METHODS: A total of 802 consecutive patients were enrolled. The Sentinel cerebral embolic protection device (Claret Medical Inc., Santa Rosa, California) was used in 34.9% (n = 280) of consecutive patients. In 65.1% (n = 522) of patients TAVR was performed in the identical setting except without cerebral embolic protection. Neurological follow-up was done within 7 days post-procedure. The primary endpoint was a composite of all-cause mortality or all-stroke according to Valve Academic Research Consortium-2 criteria within 7 days. Propensity score matching was performed to account for possible confounders.

RESULTS: Both filters of the device were successfully positioned in 280 of 305 (91.8%) consecutive patients. With use of cerebral embolic protection rate of disabling and nondisabling stroke was significantly reduced from 4.6% to 1.4% (p = 0.03; odds ratio: 0.29, 95% confidence interval: 0.10 to 0.93) in the propensity-matched population (n = 560). The primary endpoint occurred significantly less frequently, with 2.1% (n = 6 of 280) in the protected group compared with 6.8% (n = 19 of 280) in the control group (p = 0.01; odds ratio: 0.30; 95% confidence interval: 0.12 to 0.77). In multivariable analysis Society of Thoracic Surgeons score for mortality (p = 0.02) and TAVR without protection (p = 0.02) were independent predictors for the primary endpoint.

CONCLUSIONS: In patients undergoing TAVR use of a cerebral embolic protection device demonstrated a significant higher rate of stroke-free survival compared with unprotected TAVR.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app