Ocular perfusion pressure control during pars plana vitrectomy: testing a novel device.
PURPOSE: To study the efficacy of a novel device intended to control infusion pressure based on mean ocular perfusion pressure (MOPP) during pars plana vitrectomy (PPV).
METHODS: An arm blood pressure cuff connected to a vitrectomy machine calculated mean arterial pressure (MAP), while a pressure sensor close to the infusion trocar measured intraocular pressure (IOP). MOPP was calculated in real time in 36 consecutive patients undergoing PPV, who were divided into two groups. The device lowered IOP every time that calculated MOPP fell below 30 mmHg in the Control ON group (18 patients), while no action was taken in the Control OFF group (18 patients).
RESULTS: Baseline IOP and blood pressure were similar between groups. The Control ON group had significantly lower average intraoperative IOP (30.5 ± 2.1 vs. 35.9 ± 6.9 mmHg; p = 0.002) and higher MOPP (56.4 ± 5.9 vs. 49.7 ± 6.1 mmHg) than the Control OFF group. The Control ON group also spent less time at MOPP < 10 mmHg and < 30 mmHg: 0 vs. 3.40 ± 2.38 min (p < 0.001) and 9.91 ± 7.15 vs. 16.13 ± 8.12 min (p = 0.02), respectively.
CONCLUSIONS: The MOPP control device effectively maintained lower IOP and higher MOPP throughout surgery. It also helped avoid dangerous IOP peaks and MOPP dips, allowing patients to spend less time at MOPP of < 10 and < 30 mmHg.
METHODS: An arm blood pressure cuff connected to a vitrectomy machine calculated mean arterial pressure (MAP), while a pressure sensor close to the infusion trocar measured intraocular pressure (IOP). MOPP was calculated in real time in 36 consecutive patients undergoing PPV, who were divided into two groups. The device lowered IOP every time that calculated MOPP fell below 30 mmHg in the Control ON group (18 patients), while no action was taken in the Control OFF group (18 patients).
RESULTS: Baseline IOP and blood pressure were similar between groups. The Control ON group had significantly lower average intraoperative IOP (30.5 ± 2.1 vs. 35.9 ± 6.9 mmHg; p = 0.002) and higher MOPP (56.4 ± 5.9 vs. 49.7 ± 6.1 mmHg) than the Control OFF group. The Control ON group also spent less time at MOPP < 10 mmHg and < 30 mmHg: 0 vs. 3.40 ± 2.38 min (p < 0.001) and 9.91 ± 7.15 vs. 16.13 ± 8.12 min (p = 0.02), respectively.
CONCLUSIONS: The MOPP control device effectively maintained lower IOP and higher MOPP throughout surgery. It also helped avoid dangerous IOP peaks and MOPP dips, allowing patients to spend less time at MOPP of < 10 and < 30 mmHg.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app