Add like
Add dislike
Add to saved papers

Finite element analysis of customized implant in mandibular reconstruction after tumor resection with and without using customized surgical osteotomy guide.

OBJECTIVES: The aim of this work was to compare different 12 cases (3 patients * 4 cases = 12 cases) with varying gaps between implant and bone by analyzing the effect of these gaps on implant and screws using FEM.

METHODS: In each patient's case 1 using CSOG and in case 2, 3, and 4 without using CSOG tumor cutting was done. Hence in each patient zero gaps at case 1 and overcutting at case 2, 3, and 4 have obtained at different locations.

RESULTS: FEM results reveal that in each patient's case 4 (maximum gap) was more susceptible to loosening of the screws due to higher strains (37%) and implant failure due to higher stress (28%) concentration under the same loading conditions when compared with case 1 (zero gap).

CONCLUSIONS: The study reveals that mandibular reconstruction with implant placement using CSOG can significantly enhance the stability and safety of the implant.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app