Experimental Study of Right Ventricular Hemodynamics After Tricuspid Valve Replacement Therapies to Treat Tricuspid Regurgitation.

The increased understanding of right heart diseases has led to more aggressive interventions to manage functional tricuspid regurgitation (FTR). In some cases of FTR, prosthetic valve replacement is typically considered when concomitant organic components or significant geometrical distortions are involved in the pathology of the tricuspid valve. However, little is known of the performance of current devices in the right heart circulation. In this study, a novel in vitro mock circulatory system that incorporated a realistic tricuspid valve apparatus in a patient-specific silicon right ventricle (RV) was designed and fabricated. The system was calibrated to emulate severe FTR, enabling the investigation of RV hemodynamics in pre- and post-implantation of tri-leaflet tissue implant and bi-leaflet mechanical implant. 2D particle imaging velocimetry was performed to visualize flow and quantify relevant hemodynamic parameters. While our results showed all prosthetic implants improved cardiac output, these implants also subjected the RV to increased turbulence level. Our study also revealed that the implants did not create the optimal behavior of fluid transfer in the RV as we expected. Among the implants tested, tissue implant created the most dominant vortices, which persisted throughout diastole; its observed strong negative vortex could lead to increase energy expenditure due to undesired fluid direction. In contrast, both native valve and mechanical implant had both weaker vortex formation as well as more significant vortex dissipation. Interestingly, the vortex dissipation of native valve was associated with streamlined flow pattern that tended towards the pulmonary outlet, while the mechanical implant generated more regions of flow stagnation within the RV. These findings heighten the imperative to improve designs of current heart valves to be used in the right circulation.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app