Add like
Add dislike
Add to saved papers

Characterizing and eliminating errors in enhancement and subtraction artifacts in dynamic contrast-enhanced breast MRI: Chemical shift artifact of the third kind.

PURPOSE: To characterize errors in enhancement in breast dynamic contrast-enhanced (DCE) MRI studies as a function of echo time and determine the source of dark band artifacts in clinical subtraction images.

METHODS: Computer simulations, oil and water substitute (methylene chloride), as well as an American College of Radiology quality control phantom were tested. Routine clinical DCE breast MRI study was bracketed with (accelerated) in-phase DCE acquisitions in five patients.

RESULTS: Simulation results demonstrated up to -160% suppression of the expected enhancement caused by differential enhancement of fat and water. Two-dimensional gradient-recalled echo and fat-suppressed 3D GRE phantom imaging confirmed the simulation results and showed that fat suppression does not eliminate the artifact. In vivo in-phase DCE images showed increased enhancement consistent with predictions and also confirmed increased spatial blurring on in-phase 3D gradient-recalled echo images. Combined multi-dimensional partial Fourier and parallel imaging provided a time-equivalent in-phase DCE MRI acquisition.

CONCLUSION: Errors in expected enhancement occur in DCE breast MRI subtraction images because of differential enhancement of fat and water and incomplete fat signal suppression. These errors can lead to artificial suppression of enhancement as well as dark band artifacts on subtraction images. These artifacts can be eliminated with a time-equivalent in-phase fat-suppressed 3D gradient-recalled echo sequence. Understanding chemical shift artifact of the third kind, a unique artifact of artificial enhancement suppression in the presence of intravoxel fat and water signal, will aid DCE breast MRI image interpretation. In-phase acquisitions (combined with simultaneous minimum echo time or opposed-phase echoes) may facilitate qualitative, quantitative and longitudinal analysis of contrast enhancement. Magn Reson Med 79:2277-2289, 2018. © 2017 International Society for Magnetic Resonance in Medicine.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app