Evaluation of a Novel Cryoprobe for Atrial Ablation in a Chronic Ovine Model.
Annals of Thoracic Surgery 2017 September
PURPOSE: Cryoablation is used in the treatment of atrial fibrillation and other cardiac arrhythmias. This study evaluated a novel 10-cm flexible nitrous oxide cryoprobe in an ovine model of atrial ablation.
DESCRIPTION: Six sheep were anesthetized, underwent a left thoracotomy, and were placed on cardiopulmonary bypass. A left atriotomy was performed, and the cryoprobe was applied endocardially for 120 seconds at less than -40°C to 4 sites on the left atrium. The atrium was closed and the animals were allowed to recover. After 30 days, the animals were euthanized. Transmurality was evaluated in 5-mm sections of each lesion using 2,3,5-triphenyltetrazolium chloride (TTC) and Masson's trichrome staining.
EVALUATION: All animals survived. One hundred four of 106 sections (98%) were transmural by TTC; 103 of 106 (97%) sections were transmural by trichrome staining. There was no late atrial perforation, intraluminal thrombus, or thromboembolism.
CONCLUSIONS: The device reliably produced transmural lesions in a chronic ovine model. Its performance was equivalent to that of other nitrous oxide cryoablation systems.
DESCRIPTION: Six sheep were anesthetized, underwent a left thoracotomy, and were placed on cardiopulmonary bypass. A left atriotomy was performed, and the cryoprobe was applied endocardially for 120 seconds at less than -40°C to 4 sites on the left atrium. The atrium was closed and the animals were allowed to recover. After 30 days, the animals were euthanized. Transmurality was evaluated in 5-mm sections of each lesion using 2,3,5-triphenyltetrazolium chloride (TTC) and Masson's trichrome staining.
EVALUATION: All animals survived. One hundred four of 106 sections (98%) were transmural by TTC; 103 of 106 (97%) sections were transmural by trichrome staining. There was no late atrial perforation, intraluminal thrombus, or thromboembolism.
CONCLUSIONS: The device reliably produced transmural lesions in a chronic ovine model. Its performance was equivalent to that of other nitrous oxide cryoablation systems.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app