Intraoperative implantation of a mesh of directional palladium sources (CivaSheet): Dosimetry verification, clinical commissioning, dose specification, and preliminary experience.

Brachytherapy 2017 November
PURPOSE: To present the clinical commissioning of a novel 103 Pd directional brachytherapy device (CivaSheet) for intraoperative radiation therapy.

METHODS AND MATERIALS: Clinical commissioning for the CivaSheet consisted of establishing: (1) source strength calibration capabilities, (2) experimental verification of TG-43 dosimetry parameters, (3) treatment planning system validation, and (4) departmental practice for dose specification and source ordering. Experimental verification was performed in water with radiochromic film calibrated with a 37 kVp X-ray beam. Percentage difference ([measurements - calculation]/calculation) and distance to agreement (difference between film-to-source distance and distance that minimized the percentage difference) were calculated. Nomogram values (in U/100 Gy) for all configurations (up to 20 × 20 sources) were calculated for source ordering. Clinical commissioning was used on patients enrolled in an ongoing Institutional Review Board-approved protocol.

RESULTS: A source calibration procedure was established, and the treatment planning system was commissioned within standard clinical uncertainties. Percentage dose differences (distances to agreement) between measured and calculated doses were 8.6% (-0.12 mm), 0.6% (-0.01 mm), -6.4% (0.22 mm), and -10.0% (0.44 mm) at depths of 2.3, 5.1, 8.0, and 11.1 mm, respectively. All differences were within the experimental uncertainties. Nomogram values depended on sheet size and spatial extent. A value of 2.4U/100 Gy per CivaDot was found to satisfy most cases, ranging from 2.3 to 3.3U/100 Gy. Nomogram results depended on elongation of the treatment area with a higher variation observed for smaller treatment areas. Postimplantation dose evaluation was feasible.

CONCLUSIONS: Commissioning and clinical deployment of CivaSheet was feasible using BrachyVision for postoperative dose evaluation. Experimental verification confirmed that the available TG-43 dosimetry parameters are accurate for clinical use.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app