Interobserver agreement and image quality of functional cardiac ultrasound measures used in a randomised trial of delayed cord clamping in preterm infants.
OBJECTIVE: Functional cardiac ultrasound measures are used clinically and in trials for assessing the haemodynamic status of newborn infants. Superior vena cava (SVC) flow and right ventricular output (RVO) are established measures of systemic blood flow on the first postnatal day. The objective was to assess image quality and interobserver agreement of these measures in preterm infants enrolled in a randomised trial of immediate versus delayed cord clamping.
DESIGN AND SETTING: Image quality and interobserver agreement for SVC flow, RVO and ductus arteriosus (DA) size were assessed on measurements taken at 3-6, 6-12 and 20-28 hours for the first 10 infants enrolled at each of four sites (total 40). Bland-Altman plots were constructed; mean difference (bias) and limits of agreement (LOA) were calculated. Potential sources of variation were explored.
RESULTS: Quality was judged satisfactory for >97% of images. The mean difference and LOA between the observers were 5.4 mL/kg/min and -49.0 to 59.8 mL/kg/min for SVC flow, -26.6 mL/kg/min and -131.4 to 78.2 mL/kg/min for RVO, and 0 mm and -0.8 to 0.8 mm for DA diameter, respectively. The principal source of measurement error for SVC flow was diameter, and for RVO, diameter and velocity time integral. The difference between observers for both SVC and RVO was significantly associated with site.
CONCLUSION: Interobserver variability for SVC flow is consistent with that previously reported, but higher for RVO. The findings should be incorporated into clinical practice, training, accreditation and trial design.
DESIGN AND SETTING: Image quality and interobserver agreement for SVC flow, RVO and ductus arteriosus (DA) size were assessed on measurements taken at 3-6, 6-12 and 20-28 hours for the first 10 infants enrolled at each of four sites (total 40). Bland-Altman plots were constructed; mean difference (bias) and limits of agreement (LOA) were calculated. Potential sources of variation were explored.
RESULTS: Quality was judged satisfactory for >97% of images. The mean difference and LOA between the observers were 5.4 mL/kg/min and -49.0 to 59.8 mL/kg/min for SVC flow, -26.6 mL/kg/min and -131.4 to 78.2 mL/kg/min for RVO, and 0 mm and -0.8 to 0.8 mm for DA diameter, respectively. The principal source of measurement error for SVC flow was diameter, and for RVO, diameter and velocity time integral. The difference between observers for both SVC and RVO was significantly associated with site.
CONCLUSION: Interobserver variability for SVC flow is consistent with that previously reported, but higher for RVO. The findings should be incorporated into clinical practice, training, accreditation and trial design.
Full text links
Trending Papers
A Personalized Approach to the Management of Congestion in Acute Heart Failure.Heart International 2023
Potential Mechanisms of the Protective Effects of the Cardiometabolic Drugs Type-2 Sodium-Glucose Transporter Inhibitors and Glucagon-like Peptide-1 Receptor Agonists in Heart Failure.International Journal of Molecular Sciences 2024 Februrary 21
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app