Add like
Add dislike
Add to saved papers

Urea-based recirculation validation of the symmetrical palindrome catheter.

Journal of Renal Care 2017 December
BACKGROUND: High blood flow and low recirculation rates are central for adequate haemodialysis. A new symmetrical tip has been invented promising efficient haemodialysis even if the ports are reversed.

OBJECTIVE: To evaluate access recirculation of the 'palindrome' catheter and to report initial experiences in a clinical setting.

MATERIAL AND METHODS: After implantation of the new catheter in 20 patients (male: 14; female: 6; mean age 72 ± 12.2), access recirculation was evaluated using the urea-based recirculation test. After 30 minutes of haemodialysis, ultrafiltration was stopped and arterial and venous samples were taken. Afterwards, the blood flow rate was reduced to 120 ml/min. Another systemic arterial blood sample was taken 10 seconds after the blood pump was switched off.

RESULTS: All 20 interventions were performed successfully without complications. The average recirculation rate was 8.1% with a median of 2.5% ranging from 0 to 85.8%. Recirculation rates under 5% were measured in 13 patients and more than 10% recirculation were found in two patients. The median of days between catheter implantation and recirculation assessment was the day following implantation.

CONCLUSION: The new symmetrical catheter presented low recirculation rates in a clinical setting. Since there is just a single tip, fluoroscopic placement in the right atrium is facilitated.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app