We have located links that may give you full text access.
Decontamination of an Extracorporeal Membrane Oxygenator Contaminated With Mycobacterium chimaera.
Infection Control and Hospital Epidemiology 2017 October
Water samples taken from extracorporeal membrane oxygenator (ECMO) devices used at University Hospitals Birmingham yielded high total viable counts (TVCs) containing a variety of microorganisms, including M. chimaera. Disinfection resulted in the reduction of TVCs and eradication of Mycobacterium chimaera. Weekly disinfection and water sampling are required to manage the water quality in these devices. Infect Control Hosp Epidemiol 2017;38:1244-1246.