Mega Acer Kit® is more effective for warming the intravenous fluid than Ranger™ and ThermoSens® at 440 ml/h of infusion rate: an experimental performance study.

BACKGROUND: Few studies have investigated the effectiveness of intravenous fluid warmers at low and moderate flow rates below 1,000 ml/h. In this study, we compared the effectiveness of three different fluid warmers at a low flow rate (440 ml/h).

METHODS: We experimentally investigated the fluid warming performances of Mega Acer Kit® (Group M, n = 10), Ranger™ (Group R, n = 10), and ThermoSens® (Group T, n = 10) at 440 ml/h for 60 min. All devices were set at a warming temperature of 41℃ with preheating for 10 min. Intravenous fluids were then delivered through them. The fluid temperature (primary endpoint) was measured at 76 cm from the device after infusion for 60 min. The expected decrease in mean body temperature (secondary endpoint) after 5 h infusion for a 70 kg patient (ΔMBT5) was also calculated.

RESULTS: The fluid temperature (mean [95% CI]) at 76 cm from the device, 60 minutes after the infusion was higher in group M (36.01 [35.73-36.29]℃), compared to groups T (29.81 [29.38-30.24]℃) and R (29.12 [28.52-29.72]℃) (P < 0.001). The ΔMBT5 (mean [95% CI]) was significantly smaller in group M (-0.04 [-0.04 to -0.03]℃) than that in groups T (-0.27 [-0.28 to -0.29]℃; P < 0.001) and R (-0.30 [-0.32 to -0.27]℃; P < 0.001). However, none of the fluid warmers provided a constant normothermic temperature above 36.5℃.

CONCLUSIONS: Mega Acer Kit® was more effective in warming the intravenous fluid with the smallest expected change in the mean body temperature, compared to Ranger™ and ThermoSens®, at a flow rate of 440 ml/h.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app