Screw-blade fixation systems in Pauwels three femoral neck fractures: a biomechanical evaluation.

OBJECTIVES: To reduce mechanical complications after osteosynthesis of femoral neck fractures, improved fixation techniques have been developed including blade or screw-anchor devices. This biomechanical study compares different fixation systems used for treatment of unstable femoral neck fractures with evaluation of failure mode, load to failure, stiffness, femoral head rotation, femoral neck shortening and femoral head migration.

METHODS: Standardized Pauwels type 3 fractures (AO/OTA 31-B2) with comminution were created in 18 biomechanical sawbones using a custom-made sawguide. Fractures were stabilized using either SHS-Screw, SHS-Blade or Rotationally Stable Screw-Anchor (RoSA). Femurs were positioned in 25 degrees adduction and ten degrees posterior flexion and were cyclically loaded with an axial sinusoidal loading pattern of 0.5 Hz, starting with 300 N, with an increase by 300 N every 2000 cycles until bone-implant failure occurred.

RESULTS: Mean failure load for the Screw-Anchor fixation (RoSA) was 5100 N (IQR 750 N), 3900 N (IQR 75 N) for SHS-Blade and 3000 N (IQR 675 N; p = 0.002) for SHS-Screw. For SHS-Screw and SHS-Blade we observed fracture displacement with consecutive fracture collapse as the main reason for failure, whereas RoSA mainly showed a cut-out under high loadings. Mean stiffness at 1800 N was 826 (IQR 431) N/mm for SHS-Screw, 1328 (IQR 441) N/mm for SHS-Blade and 1953 (IQR 617) N/mm for RoSA (p = 0.003). With a load of 1800 N (SHS-Screw 12° vs. SHS-Blade 7° vs. RoSA 2°; p = 0.003) and with 2700 N (24° vs. 15° vs. 3°; p = 0.002) the RoSA implants demonstrated a higher rotational stability and had the lowest femoral neck shortening (p = 0.002), compared with the SHS groups. At the 2700 N load point, RoSA systems showed a lower axial (p = 0.019) and cranial (p = 0.031) femoral head migration compared to the SHS-Screw.

CONCLUSIONS: In our study, the new Screw-Anchor fixation (RoSA) was superior to the comparable SHS implants regarding rotational stability and femoral neck shortening. Failure load, stiffness, femoral head migration, and resistance to fracture displacement were in RoSA implants higher than in SHS-Screws, but without significance in comparison to SHS-Blades.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app