Point-of-care measurement of fetal blood lactate - Time to trust a new device.
Australian & New Zealand Journal of Obstetrics & Gynaecology 2018 Februrary
BACKGROUND: Point-of-care lactate devices are used worldwide for intrapartum decision making. Current practice is often based on Lactate Pro (Arkray) but its imminent product discontinuation necessitates determination of an optimal replacement device.
AIMS: To evaluate the performance of Lactate Pro and two other point-of-care devices, Lactate Pro 2 (Arkray) and StatStrip (Nova Biomedical), and to derive scalp lactate cut-offs equivalent to the current intervention trigger of >4.8 mmol/L.
MATERIALS AND METHODS: Paired umbilical cord arterial and venous blood samples from 109 births were tested on the three point-of-care products (two devices each), cross-compared with the reference method blood gas analyser.
RESULTS: All brands deviate from the blood gas analyser, with Lactate Pro and StatStrip results consistently lower and Lactate Pro 2 consistently higher. Standard deviation from the blood gas analyser was smallest for StatStrip (0.78 mmol/L, cord artery), and largest for Lactate Pro 2 (1.03 mmol/L, cord artery). Within-brand variation exists and is similar for all brands (mean absolute difference on cord artery 0.23-0.30 mmol/L). Equivalent values to the 4.8 mmol/L intervention threshold based on Lactate Pro are 4.9-5.0 mmol/L for StatStrip and 5.3-5.9 mmol/L for Lactate Pro 2, calculated by receiver-operating characteristic analysis.
CONCLUSIONS: StatStrip appears superior to Lactate Pro 2 to replace the original Lactate Pro. Using StatStrip, the 4.8 mmol/L intervention threshold equivalent was 4.9-5.0 mmol/L. The variation in accuracy of point-of-care lactate devices may exceed the small increments (eg <4.2 mmol/L vs >4.8 mmol/L) that guide obstetric decisions.
AIMS: To evaluate the performance of Lactate Pro and two other point-of-care devices, Lactate Pro 2 (Arkray) and StatStrip (Nova Biomedical), and to derive scalp lactate cut-offs equivalent to the current intervention trigger of >4.8 mmol/L.
MATERIALS AND METHODS: Paired umbilical cord arterial and venous blood samples from 109 births were tested on the three point-of-care products (two devices each), cross-compared with the reference method blood gas analyser.
RESULTS: All brands deviate from the blood gas analyser, with Lactate Pro and StatStrip results consistently lower and Lactate Pro 2 consistently higher. Standard deviation from the blood gas analyser was smallest for StatStrip (0.78 mmol/L, cord artery), and largest for Lactate Pro 2 (1.03 mmol/L, cord artery). Within-brand variation exists and is similar for all brands (mean absolute difference on cord artery 0.23-0.30 mmol/L). Equivalent values to the 4.8 mmol/L intervention threshold based on Lactate Pro are 4.9-5.0 mmol/L for StatStrip and 5.3-5.9 mmol/L for Lactate Pro 2, calculated by receiver-operating characteristic analysis.
CONCLUSIONS: StatStrip appears superior to Lactate Pro 2 to replace the original Lactate Pro. Using StatStrip, the 4.8 mmol/L intervention threshold equivalent was 4.9-5.0 mmol/L. The variation in accuracy of point-of-care lactate devices may exceed the small increments (eg <4.2 mmol/L vs >4.8 mmol/L) that guide obstetric decisions.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app