Accuracy of 3D fluoro-navigated anterior transpedicular screws in the subaxial cervical spine: an experimental study on human specimens.
European Spine Journal 2017 November
INTRODUCTION: Pedicle screw stabilization, the standard technique in the thoracic and lumbar spine, is increasingly used in the cervical spine. Initial studies on the use of anterior pedicle screws (ATPS) in the cervical spine have been recently published. ATPS use has theoretical advantages over posterior stabilization. We have already established a 3D-fluoroscopy navigation setup in a study of artificial bones. The aim of the current study was to evaluate the positioning quality/accuracy of ATPS introduced to human specimens.
METHODS: 36 cannulated screws (3.5 mm) were implanted anteriorly into the C3-C7 segments of four spines (unfixed, frozen, cadaveric specimens) using a 3D-fluoroscopy navigation system. Placement accuracy was evaluated using a recently published classification on postoperative CT scans. Grade 1 is perfect position with pedicle wall perforation <1 mm, grade 2 is perforation <2 mm, etc., and finally grade 5 is cortical perforation of >4 mm and/or transverse foramen entry.
RESULTS: 36 anterior pedicle screws were inserted into four human cervical spine specimens. Of these, seven screws were introduced to C3, five to C4 and eight each to C5, C6, and C7. Classified with the modified G&R, 21 of 36 (58.3%) were grade 1. Ten screws (27.8%) were grade 2. Grade 4 was assessed for two screws and grade 5 for three. Customary "good" positioning, combining grades 1 and 2, was thus found in 86.1%. Five screws (13.9%) did not meet this criterion (grade ≥3).
CONCLUSIONS: With 86.1% of good positioning (grade 2 or better), a 3D-fluoroscopy navigation of ATPS screws into human c-spine specimens achieved a satisfying results. These are at least comparable to results presented in the literature for posteriorly introduced subaxial pedicle screws.
METHODS: 36 cannulated screws (3.5 mm) were implanted anteriorly into the C3-C7 segments of four spines (unfixed, frozen, cadaveric specimens) using a 3D-fluoroscopy navigation system. Placement accuracy was evaluated using a recently published classification on postoperative CT scans. Grade 1 is perfect position with pedicle wall perforation <1 mm, grade 2 is perforation <2 mm, etc., and finally grade 5 is cortical perforation of >4 mm and/or transverse foramen entry.
RESULTS: 36 anterior pedicle screws were inserted into four human cervical spine specimens. Of these, seven screws were introduced to C3, five to C4 and eight each to C5, C6, and C7. Classified with the modified G&R, 21 of 36 (58.3%) were grade 1. Ten screws (27.8%) were grade 2. Grade 4 was assessed for two screws and grade 5 for three. Customary "good" positioning, combining grades 1 and 2, was thus found in 86.1%. Five screws (13.9%) did not meet this criterion (grade ≥3).
CONCLUSIONS: With 86.1% of good positioning (grade 2 or better), a 3D-fluoroscopy navigation of ATPS screws into human c-spine specimens achieved a satisfying results. These are at least comparable to results presented in the literature for posteriorly introduced subaxial pedicle screws.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app