PURPOSE: Although capnography is considered the gold standard for monitoring of ventilation, it may not work accurately in some situations. We compared the performance of two non-invasive continuous respiratory rate (RR) monitors that are alternatives for the detection of respiratory depression.

METHODS: Fifty healthy volunteers ≥20 years old were enrolled in this study. After monitoring of the volunteers was started by the Masimo Radical-7® and Nellcor™ PM1000N, they breathed at the rate of 12 breaths/min for 3 min and then stopped breathing. As soon as the apnea alarm of either monitor went off, breathing was resumed at the same rate. This entire procedure was repeated three times. The data collected every 30 s included the RR, pulse rate (PR) and oxygen saturation (SpO2 ). The times of alarm on, alarm off and reappearance of RR on the monitor screens were also recorded.

RESULTS: The biases of the RR, PR and SpO2 measurements from the two monitors were 0.5, 0.2 and -0.4, respectively. Of 143 procedures that stopped breathing for more than 30 s, 114 and 15 alarms were shown by the Masimo Radical-7® and Nellcor™ PM-1000N monitors, respectively. Most alarms of the Nellcor™ PM1000N followed from SpO2 <90%. Conversely, most alarms of the Masimo Radical-7® were caused by RR <10 breaths/min. Times to alarm on, alarm off and display of RR measured by the Masimo Radical-7® monitor were significantly shorter than the Nellcor™ PM-1000N monitor.

CONCLUSIONS: The Masimo Radical-7® monitor provides better detection of apnea in volunteers than the Nellcor™ PM-1000N.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app