Adapting VMAT plans optimized for an HD120 MLC for delivery with a Millennium MLC.

Linac downtime invariably impacts delivery of patients' scheduled treatments. Transferring a patient's treatment to an available linac is a common practice. Transferring a Volumetric Modulated Arc Therapy (VMAT) plan from a linac equipped with a standard-definition MLC to one equipped with a higher definition MLC is practical and routine in clinics with multiple MLC-equipped linacs. However, the reverse transfer presents a challenge because the high-definition MLC aperture shapes must be adapted for delivery with the lower definition device. We have developed an efficient method to adapt VMAT plans originally designed for a high-definition MLC to a standard-definition MLC. We present the dosimetric results of our adaptation method for head-and-neck, brain, lung, and prostate VMAT plans. The delivery of the adapted plans was verified using standard phantom measurements.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app