We have located links that may give you full text access.
Recording characteristics of electrical impedance myography needle electrodes.
Physiological Measurement 2017 August 22
OBJECTIVE: Neurologists and physiatrists need improved tools for the evaluation of skeletal muscle condition. Here we evaluate needle electrical impedance myography (EIM), a new minimally invasive approach to determine muscle status that could ultimately become a bedside tool for the assessment of neuromuscular disorders.
APPROACH: We design and study the recording characteristics of tetrapolar EIM needle electrodes combining theory and finite-element model simulations. We then use these results to build and pilot in vivo an EIM needle electrode in the rat gastrocnemius muscle ([Formula: see text]). The dielectric properties of muscle are reported (mean ± standard deviation).
RESULTS: The numerical simulations show that the contribution of subcutaneous fat and muscle tissues to needle EIM data is <3% and >97%, respectively, and the sensed volume is [Formula: see text] cm3 . Apparent resistivity [Formula: see text] [Formula: see text] cm and relative permittivity [Formula: see text] (dimensionless) measured at 10 kHz are in good agreement with in vivo dielectric properties reported in the literature.
SIGNIFICANCE: The results presented show the feasibility of measuring muscle impedivity in vivo using a needle electrode from 10 kHz to 1 MHz. The development of needle EIM technology can open up a new field of study in electrodiagnostic medicine, with potential applications to both disease diagnosis and biomarker assessment of therapy.
APPROACH: We design and study the recording characteristics of tetrapolar EIM needle electrodes combining theory and finite-element model simulations. We then use these results to build and pilot in vivo an EIM needle electrode in the rat gastrocnemius muscle ([Formula: see text]). The dielectric properties of muscle are reported (mean ± standard deviation).
RESULTS: The numerical simulations show that the contribution of subcutaneous fat and muscle tissues to needle EIM data is <3% and >97%, respectively, and the sensed volume is [Formula: see text] cm3 . Apparent resistivity [Formula: see text] [Formula: see text] cm and relative permittivity [Formula: see text] (dimensionless) measured at 10 kHz are in good agreement with in vivo dielectric properties reported in the literature.
SIGNIFICANCE: The results presented show the feasibility of measuring muscle impedivity in vivo using a needle electrode from 10 kHz to 1 MHz. The development of needle EIM technology can open up a new field of study in electrodiagnostic medicine, with potential applications to both disease diagnosis and biomarker assessment of therapy.
Full text links
Related Resources
Trending Papers
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app