Response of synthetic diamond detectors in proton, carbon, and oxygen ion beams.

Medical Physics 2017 October
PURPOSE: In this work, the LET-dependence of the response of synthetic diamond detectors is investigated in different particle beams.

METHOD: Measurements were performed in three nonmodulated particle beams (proton, carbon, and oxygen). The response of five synthetic diamond detectors was compared to the response of a Markus or an Advanced Markus ionization chamber. The synthetic diamond detectors were used with their axis parallel to the beam axis and without any bias voltage. A high bias voltage was applied to the ionization chambers, to minimize ion recombination, for which no correction is applied (+300 V and +400 V were applied to the Markus and Advanced Markus ionization chambers respectively).

RESULTS: The ratio between the normalized response of the synthetic diamond detectors and the normalized response of the ionization chamber shows an under-response of the synthetic diamond detectors in carbon and oxygen ion beams. No under-response of the synthetic diamond detectors is observed in protons. For each beam, combining results obtained for the five synthetic diamond detectors and considering the uncertainties, a linear fit of the ratio between the normalized response of the synthetic diamond detectors and the normalized response of the ionization chamber is determined. The response of the synthetic diamond detectors can be described as a function of LET as (-6.22E-4 ± 3.17E-3) • LET + (0.99 ± 0.01) in proton beam, (-2.51E-4 ± 1.18E-4) • LET + (1.01 ± 0.01) in carbon ion beam and (-2.77E-4 ± 0.56E-4) • LET + (1.03 ± 0.01) in oxygen ion beam. Combining results obtained in carbon and oxygen ion beams, a LET dependence of about 0.026% (±0.013%) per keV/μm is estimated.

CONCLUSIONS: Due to the high LET value, a LET dependence of the response of the synthetic diamond detector was observed in the case of carbon and oxygen beams. The effect was found to be negligible in proton beams, due to the low LET value. The under-response of the synthetic diamond detector may result from the recombination of electron/hole in the thin synthetic diamond layer, due to the high LET-values. More investigations are required to confirm this assumption.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app