Evaluation of BiliCare™ transcutaneous bilirubin device in Japanese newborns.
BACKGROUND: Non-invasive transcutaneous bilirubin (TcB) monitoring has been widely used to screen for hyperbilirubinemia. TcB measured using the recently developed BiliCare™ system, however, has not been fully evaluated.
METHODS: One hundred and seven TcB measurements were obtained from 82 Japanese newborns ≥35 weeks' gestational age within 2 weeks after birth. Measurements were taken at the scaphoid fossa, conchal cavity, and lobe of the ear using BiliCare. BiliCare TcB were compared with total serum bilirubin (TB) and TcB obtained using another bilirubinometer (JM-105™).
RESULTS: Transcutaneous bilirubin measured at all three sites significantly correlated with TB (r = 0.91, 0.93, and 0.93 at the scaphoid fossa, conchal cavity, and lobe, respectively). The mean differences were 0.1, -0.3, and 3.6 at the scaphoid fossa, conchal cavity, and lobe, respectively. BiliCare TcB at the scaphoid fossa significantly correlated with that using the JM-105 (r = 0.91). The mean difference was 0.0. BiliCare, however, produced a significantly higher and lower TcB than the JM-105 for TB <7 and ≥15 mg/dL, respectively.
CONCLUSIONS: Transcutaneous bilirubin measurements taken at the scaphoid fossa or conchal cavity using BiliCare were more reliable than those at the earlobe. BiliCare TcB differed from those of the JM-105, for TB <7 or ≥15 mg/dL.
METHODS: One hundred and seven TcB measurements were obtained from 82 Japanese newborns ≥35 weeks' gestational age within 2 weeks after birth. Measurements were taken at the scaphoid fossa, conchal cavity, and lobe of the ear using BiliCare. BiliCare TcB were compared with total serum bilirubin (TB) and TcB obtained using another bilirubinometer (JM-105™).
RESULTS: Transcutaneous bilirubin measured at all three sites significantly correlated with TB (r = 0.91, 0.93, and 0.93 at the scaphoid fossa, conchal cavity, and lobe, respectively). The mean differences were 0.1, -0.3, and 3.6 at the scaphoid fossa, conchal cavity, and lobe, respectively. BiliCare TcB at the scaphoid fossa significantly correlated with that using the JM-105 (r = 0.91). The mean difference was 0.0. BiliCare, however, produced a significantly higher and lower TcB than the JM-105 for TB <7 and ≥15 mg/dL, respectively.
CONCLUSIONS: Transcutaneous bilirubin measurements taken at the scaphoid fossa or conchal cavity using BiliCare were more reliable than those at the earlobe. BiliCare TcB differed from those of the JM-105, for TB <7 or ≥15 mg/dL.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app