Navigation accuracy after automatic- and hybrid-surface registration in sinus and skull base surgery.
PloS One 2017
OBJECTIVE: Computer-aided-surgery in ENT surgery is mainly used for sinus surgery but navigation accuracy still reaches its limits for skull base procedures. Knowledge of navigation accuracy in distinct anatomical regions is therefore mandatory. This study examined whether navigation accuracy can be improved in specific anatomical localizations by using hybrid registration technique.
STUDY DESIGN: Experimental phantom study.
SETTING: Operating room.
SUBJECTS AND METHODS: The gold standard of screw registration was compared with automatic LED-mask-registration alone, and in combination with additional surface matching. 3D-printer-based skull models with individual fabricated silicone skin were used for the experiments. Overall navigation accuracy considering 26 target fiducials distributed over each skull was measured as well as the accuracy on selected anatomic localizations.
RESULTS: Overall navigation accuracy was <1.0 mm in all cases, showing the significantly lowest values after screw registration (0.66 ± 0.08 mm), followed by hybrid registration (0.83± 0.08 mm), and sole mask registration (0.92 ± 0.13 mm).On selected anatomic localizations screw registration was significantly superior on the sphenoid sinus and on the internal auditory canal. However, mask registration showed significantly better accuracy results on the midface. Navigation accuracy on skull base localizations could be significantly improved by the combination of mask registration and additional surface matching.
CONCLUSION: Overall navigation accuracy gives no sufficient information regarding navigation accuracy in a distinct anatomic area. The non-invasive LED-mask-registration proved to be an alternative in clinical routine showing best accuracy results on the midface. For challenging skull base procedures a hybrid registration technique is recommendable which improves navigation accuracy significantly in this operating field. Invasive registration procedures are reserved for selected challenging skull base operations where the required high precision warrants the invasiveness.
STUDY DESIGN: Experimental phantom study.
SETTING: Operating room.
SUBJECTS AND METHODS: The gold standard of screw registration was compared with automatic LED-mask-registration alone, and in combination with additional surface matching. 3D-printer-based skull models with individual fabricated silicone skin were used for the experiments. Overall navigation accuracy considering 26 target fiducials distributed over each skull was measured as well as the accuracy on selected anatomic localizations.
RESULTS: Overall navigation accuracy was <1.0 mm in all cases, showing the significantly lowest values after screw registration (0.66 ± 0.08 mm), followed by hybrid registration (0.83± 0.08 mm), and sole mask registration (0.92 ± 0.13 mm).On selected anatomic localizations screw registration was significantly superior on the sphenoid sinus and on the internal auditory canal. However, mask registration showed significantly better accuracy results on the midface. Navigation accuracy on skull base localizations could be significantly improved by the combination of mask registration and additional surface matching.
CONCLUSION: Overall navigation accuracy gives no sufficient information regarding navigation accuracy in a distinct anatomic area. The non-invasive LED-mask-registration proved to be an alternative in clinical routine showing best accuracy results on the midface. For challenging skull base procedures a hybrid registration technique is recommendable which improves navigation accuracy significantly in this operating field. Invasive registration procedures are reserved for selected challenging skull base operations where the required high precision warrants the invasiveness.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
Perioperative echocardiographic strain analysis: what anesthesiologists should know.Canadian Journal of Anaesthesia 2024 April 11
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app