Reverse total shoulder glenoid baseplate stability with superior glenoid bone loss.
Journal of Shoulder and Elbow Surgery 2017 October
BACKGROUND: Superior wear of the glenoid bone is common in patients with rotator cuff arthropathy. This can become a treatment challenge for patients who require shoulder arthroplasty. In reverse shoulder arthroplasty (RSA), glenoid bone loss may affect the stability of baseplate fixation. The primary purpose of this biomechanical laboratory study was to assess the initial fixation stability of RSA glenosphere baseplates in the presence of variable amounts of superior glenoid bone loss.
MATERIALS AND METHODS: High-density solid rigid polyurethane foam (30 pounds/cubic foot) was machined to model the glenoid with variable superior defects that provided different levels of support (100%, 90%, 75%, and 50%) for the glenosphere baseplate. The samples were cyclically loaded (0-750 N at 1 Hz for 5000 cycles) at a 60° glenohumeral angle. The micromotion and migration of the baseplate were calculated from displacement data captured during the loading tests with an array of 3 linear variable differential transformers mounted around the baseplate.
RESULTS: Micromotion was significantly greater in samples with 50% defects compared with those with smaller defects. Migration was significantly greater after testing for all defect sizes.
CONCLUSIONS: Initial fixation of RSA glenosphere baseplates was significantly reduced in models with 50% bone loss on the superior edge compared with models with less bone loss in this high-density bone foam model.
MATERIALS AND METHODS: High-density solid rigid polyurethane foam (30 pounds/cubic foot) was machined to model the glenoid with variable superior defects that provided different levels of support (100%, 90%, 75%, and 50%) for the glenosphere baseplate. The samples were cyclically loaded (0-750 N at 1 Hz for 5000 cycles) at a 60° glenohumeral angle. The micromotion and migration of the baseplate were calculated from displacement data captured during the loading tests with an array of 3 linear variable differential transformers mounted around the baseplate.
RESULTS: Micromotion was significantly greater in samples with 50% defects compared with those with smaller defects. Migration was significantly greater after testing for all defect sizes.
CONCLUSIONS: Initial fixation of RSA glenosphere baseplates was significantly reduced in models with 50% bone loss on the superior edge compared with models with less bone loss in this high-density bone foam model.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app