Image improvement method for positron emission mammography.
Physica Medica : PM 2017 July
PURPOSE: To evaluate in clinical use a rapidly converging, efficient iterative deconvolution algorithm (RSEMD) for improving the quantitative accuracy of previously reconstructed breast images by a commercial positron emission mammography (PEM) scanner.
MATERIALS AND METHODS: The RSEMD method was tested on imaging data from clinical Naviscan Flex Solo II PEM scanner. This method was applied to anthropomorphic like breast phantom data and patient breast images previously reconstructed with Naviscan software to determine improvements in image resolution, signal to noise ratio (SNR) and contrast to noise ratio (CNR).
RESULTS: In all of the patients' breast studies the improved images proved to have higher resolution, contrast and lower noise as compared with images reconstructed by conventional methods. In general, the values of CNR reached a plateau at an average of 6 iterations with an average improvement factor of about 2 for post-reconstructed Flex Solo II PEM images. Improvements in image resolution after the application of RSEMD have also been demonstrated.
CONCLUSIONS: A rapidly converging, iterative deconvolution algorithm with a resolution subsets-based approach (RSEMD) that operates on patient DICOM images has been used for quantitative improvement in breast imaging. The RSEMD method can be applied to PEM images to enhance the resolution and contrast in cancer diagnosis to monitor the tumor progression at the earliest stages.
MATERIALS AND METHODS: The RSEMD method was tested on imaging data from clinical Naviscan Flex Solo II PEM scanner. This method was applied to anthropomorphic like breast phantom data and patient breast images previously reconstructed with Naviscan software to determine improvements in image resolution, signal to noise ratio (SNR) and contrast to noise ratio (CNR).
RESULTS: In all of the patients' breast studies the improved images proved to have higher resolution, contrast and lower noise as compared with images reconstructed by conventional methods. In general, the values of CNR reached a plateau at an average of 6 iterations with an average improvement factor of about 2 for post-reconstructed Flex Solo II PEM images. Improvements in image resolution after the application of RSEMD have also been demonstrated.
CONCLUSIONS: A rapidly converging, iterative deconvolution algorithm with a resolution subsets-based approach (RSEMD) that operates on patient DICOM images has been used for quantitative improvement in breast imaging. The RSEMD method can be applied to PEM images to enhance the resolution and contrast in cancer diagnosis to monitor the tumor progression at the earliest stages.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app