Performance of automated digital cell imaging analyzer Sysmex DI-60.

BACKGROUND: The Sysmex DI-60 system (DI-60, Sysmex, Kobe, Japan) is a new automated digital cell imaging analyzer. We explored the performance of DI-60 in comparison with Sysmex XN analyzer (XN, Sysmex) and manual count.

METHODS: In a total of 276 samples (176 abnormal and 100 normal samples), white blood cell (WBC) differentials, red blood cell (RBC) classification and platelet (PLT) estimation by DI-60 were compared with the results by XN and/or manual count. RBC morphology between pre-classification and verification was compared according to the ICSH grading criteria. The manual count was performed according to the Clinical and Laboratory Standards Institute guidelines (H20-A2).

RESULTS: The overall concordance between DI-60 and manual count for WBCs was 86.0%. The agreement between DI-60 pre-classification and verification was excellent (weighted κ=0.963) for WBC five-part differentials. The correlation with manual count was very strong for neutrophils (r=0.955), lymphocytes (r=0.871), immature granulocytes (r=0.820), and blasts (r=0.879). RBC grading showed notable differences between DI-60 and manual counting on the basis of the ICSH grading criteria. Platelet count by DI-60 highly correlated with that by XN (r=0.945). However, DI-60 underestimated platelet counts in samples with marked thrombocytosis.

CONCLUSIONS: The performance of DI-60 for WBC differential, RBC classification, and platelet estimation seems to be acceptable even in abnormal samples with improvement after verification. DI-60 would help optimize the workflow in hematology laboratory with reduced manual workload.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app