Evaluation of the Sysmex XN-550, a Novel Compact Haematology analyser from the XN-L ® series, compared to the XN-20 system.
International Journal of Laboratory Hematology 2017 December
INTRODUCTION: The XN-550 is a new, automated, compact, haematology analyser designed to generate a full blood count with a standard five-part white blood cell differential and an immature granulocyte count, as well as an optional reticulocyte and optical platelet (PLT) counts. The aim of the study was to evaluate the performance of the XN-550 and compare it to the established XN-20 system.
METHODS: We evaluated the basic parameter and special measurement channels of the XN-550, using the XN-20 (which has a similar operating system), as a reference analyser. Precision, carry-over and throughput evaluations were performed. In addition, a total of 202 samples including normal controls and various pathological samples were studied for comparability.
RESULTS: Good correlations with the reference analyser were obtained for all parameters except basophils. The XN-550 offers impedance and optical PLT counts and the latter showed a better correlation and less scatter than the impedance count and was comparable to the XN-20 fluorescent count at PLT counts ≤40×109 /L. Precision was good, and no significant carry-over was detected.
CONCLUSIONS: The XN-550 was simple and easy to use, while maintaining the good diagnostic sensitivity seen with high-range systems such as the XN-20, making this compact device suitable for near-patient services and smaller satellite laboratories.
METHODS: We evaluated the basic parameter and special measurement channels of the XN-550, using the XN-20 (which has a similar operating system), as a reference analyser. Precision, carry-over and throughput evaluations were performed. In addition, a total of 202 samples including normal controls and various pathological samples were studied for comparability.
RESULTS: Good correlations with the reference analyser were obtained for all parameters except basophils. The XN-550 offers impedance and optical PLT counts and the latter showed a better correlation and less scatter than the impedance count and was comparable to the XN-20 fluorescent count at PLT counts ≤40×109 /L. Precision was good, and no significant carry-over was detected.
CONCLUSIONS: The XN-550 was simple and easy to use, while maintaining the good diagnostic sensitivity seen with high-range systems such as the XN-20, making this compact device suitable for near-patient services and smaller satellite laboratories.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app