We have located links that may give you full text access.
Effect of hand hygiene and glove use on cleanliness of reusable surgical instruments.
Journal of Hospital Infection 2017 December
BACKGROUND: During functionality testing and packaging of reusable surgical instruments (RSI) for sterilization, instruments are frequently touched. There is a lack of standards relating to hand hygiene frequency and use of gloves in the sterilizing service unit packing area.
AIM: To determine the effect of hand hygiene and glove use on maintenance of RSI cleanliness.
METHODS: Following manual and automated cleaning, Halsted-mosquito forceps were assessed for adenosine triphosphate (ATP), protein and microbial contamination after handling with gloved and ungloved but washed hands using an ATP surface swab test, bicinchoninic acid assay, and standard culture plate/broth, respectively. Gram's stain was used to classify the isolates. RSI contamination was assessed immediately following and 1, 2, and 4 h after washing hands.
FINDINGS: Packing instruments with hands that had been unwashed for 2 or 4 h resulted in a significant increase in contaminating ATP when compared with all other treatment groups (P < 0.05). There was a significant correlation between the time since washing hands, the amount of ATP (r = 0.93; P ≤ 0.001), and the microbial load (r = 0.83; P ≤ 0.001) contaminating the forceps, where the longer the time the hands remained unwashed the higher the contamination. Significantly more contaminating protein was found on forceps handled with ungloved hands that had not been washed for 2 or 4 h (P < 0.001).
CONCLUSION: Critical RSI inspection, assembling, lubricating and packing should be performed using either gloves or within 1 h of washing hands.
AIM: To determine the effect of hand hygiene and glove use on maintenance of RSI cleanliness.
METHODS: Following manual and automated cleaning, Halsted-mosquito forceps were assessed for adenosine triphosphate (ATP), protein and microbial contamination after handling with gloved and ungloved but washed hands using an ATP surface swab test, bicinchoninic acid assay, and standard culture plate/broth, respectively. Gram's stain was used to classify the isolates. RSI contamination was assessed immediately following and 1, 2, and 4 h after washing hands.
FINDINGS: Packing instruments with hands that had been unwashed for 2 or 4 h resulted in a significant increase in contaminating ATP when compared with all other treatment groups (P < 0.05). There was a significant correlation between the time since washing hands, the amount of ATP (r = 0.93; P ≤ 0.001), and the microbial load (r = 0.83; P ≤ 0.001) contaminating the forceps, where the longer the time the hands remained unwashed the higher the contamination. Significantly more contaminating protein was found on forceps handled with ungloved hands that had not been washed for 2 or 4 h (P < 0.001).
CONCLUSION: Critical RSI inspection, assembling, lubricating and packing should be performed using either gloves or within 1 h of washing hands.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app