Comparison of central corneal thickness with four noncontact devices: An agreement analysis of swept-source technology.
Indian Journal of Ophthalmology 2017 June
PURPOSE: The purpose of this study was to compare the central corneal thickness (CCT) measurements of four noncontact devices in healthy eyes.
MATERIALS AND METHODS: In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-domain OCT (SD-OCT) device with an anterior segment module (Spectralis), and noncontact pachymetry (NCP) device (Topcon TRK-2P). Agreement among the devices was analyzed using mean differences (i.e., bias) and Bland-Altman analysis with 95% limits of agreement (LoA).
RESULTS: Mean CCT measurements were 537.5 ± 47.5 μm for SS-OCT optical biometer, 532.3 ± 43.8 μm for Scheimpflug system, 521.3 ± 44.7 μm for SD-OCT device, and 518.0 ± 43.1 μm for NCP (P < 0.001). The SD-OCT device and NCP showed the closest agreement, with a bias of 2.6 μm (95% LoA, -3.6-8.8 μm), whereas the SS-OCT optical biometer and NCP showed the least agreement, with a bias of 18.7 μm (95% LoA, -2.1-39.5 μm). Bias was 16.1 μm (95% LoA, -3.1-35.3 μm) for SS-OCT optical biometer and SD-OCT, 5.1 μm (95% LoA, -6.8-17.0 μm) for SS-OCT optical biometer and Scheimpflug system, 10.9 μm (95% LoA, -15.1-36.9 μm) for SD-OCT device and Scheimpflug system, and 13.6 μm (95% LoA, -5-32.2 μm) for Scheimpflug system and NCP.
CONCLUSIONS: SS-OCT optical biometer overestimates CCT measurements compared to Scheimpflug system, SD-OCT device, and NCP. Given mean differences and range variations in CCT measurements between devices, SS-OCT optical biometer and Scheimpflug system are interchangeable as are SD-OCT and NCP.
MATERIALS AND METHODS: In a sample of 45 healthy controls, CCT was measured using an optical biometer (IOLMaster 700) based on swept-source optical coherence tomography (SS-OCT), high-resolution rotating Scheimpflug camera system (Pentacam HR), spectral-domain OCT (SD-OCT) device with an anterior segment module (Spectralis), and noncontact pachymetry (NCP) device (Topcon TRK-2P). Agreement among the devices was analyzed using mean differences (i.e., bias) and Bland-Altman analysis with 95% limits of agreement (LoA).
RESULTS: Mean CCT measurements were 537.5 ± 47.5 μm for SS-OCT optical biometer, 532.3 ± 43.8 μm for Scheimpflug system, 521.3 ± 44.7 μm for SD-OCT device, and 518.0 ± 43.1 μm for NCP (P < 0.001). The SD-OCT device and NCP showed the closest agreement, with a bias of 2.6 μm (95% LoA, -3.6-8.8 μm), whereas the SS-OCT optical biometer and NCP showed the least agreement, with a bias of 18.7 μm (95% LoA, -2.1-39.5 μm). Bias was 16.1 μm (95% LoA, -3.1-35.3 μm) for SS-OCT optical biometer and SD-OCT, 5.1 μm (95% LoA, -6.8-17.0 μm) for SS-OCT optical biometer and Scheimpflug system, 10.9 μm (95% LoA, -15.1-36.9 μm) for SD-OCT device and Scheimpflug system, and 13.6 μm (95% LoA, -5-32.2 μm) for Scheimpflug system and NCP.
CONCLUSIONS: SS-OCT optical biometer overestimates CCT measurements compared to Scheimpflug system, SD-OCT device, and NCP. Given mean differences and range variations in CCT measurements between devices, SS-OCT optical biometer and Scheimpflug system are interchangeable as are SD-OCT and NCP.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app