[Clinical Study of Rs-EPI DWI to Improve Imaging Quality and Assessment on Intracranial Mass Lesions].
Sichuan da Xue Xue Bao. Yi Xue Ban = Journal of Sichuan University. Medical Science Edition 2017 May
OBJECTIVES: To explore the clinical value of readout-segmented echo planar imaging (Rs-EPI) sequence in the assessment of intracranial mass lesions compared to the standard single-shot EPI (Ss-EPI) sequence.
METHODS: We included 21 patients with intracranial mass lesions who underwent both Ss-EPI diffusion weighted imaging (DWI) and Rs-EPI DWI at 3.0T MR scanner with a twenty-channel head-neck coil. The quality of images was assessed by two experienced radiologists independently. The differences in image quality between two sequences were analyzed using Wilcoxon signed-rank test. Inter-observer agreements were analyzed using interclass correlation coefficient (ICC) and Kappa test.
RESULTS: All objectives were completed on 3.0T MR. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in Rs-EPI DWI were higher than those in Ss-EPI DWI (130.46±49.10 vs. 71.58±30.43, P=0.000; 33.22±18.86 vs. 17.92±18.72, P=0.003). The scores of overall image quality, ghost artifact where next to the paranasal sinuses, mastoid air cells and frontal sinus of Rs-EPI DWI were significantly higher than those of Ss-EPI DWI. Meanwhile, the geometric distortion of anatomical structures of Rs-EPI DWI were significantly lower compared to Ss-EPI DWI sequence (0.016±0.021 vs. 0.037±0.069, P=0.00). The inter-reader and intra-reader agreements for the assessment of qualitative parameters were good [0.74≤Kappa value or ICC≤0.92].
CONCLUSIONS: Rs-EPI DWI sequence is a potential technique to improve the imaging quality in the diagnosis of intracranial mass lesions.
METHODS: We included 21 patients with intracranial mass lesions who underwent both Ss-EPI diffusion weighted imaging (DWI) and Rs-EPI DWI at 3.0T MR scanner with a twenty-channel head-neck coil. The quality of images was assessed by two experienced radiologists independently. The differences in image quality between two sequences were analyzed using Wilcoxon signed-rank test. Inter-observer agreements were analyzed using interclass correlation coefficient (ICC) and Kappa test.
RESULTS: All objectives were completed on 3.0T MR. The signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) in Rs-EPI DWI were higher than those in Ss-EPI DWI (130.46±49.10 vs. 71.58±30.43, P=0.000; 33.22±18.86 vs. 17.92±18.72, P=0.003). The scores of overall image quality, ghost artifact where next to the paranasal sinuses, mastoid air cells and frontal sinus of Rs-EPI DWI were significantly higher than those of Ss-EPI DWI. Meanwhile, the geometric distortion of anatomical structures of Rs-EPI DWI were significantly lower compared to Ss-EPI DWI sequence (0.016±0.021 vs. 0.037±0.069, P=0.00). The inter-reader and intra-reader agreements for the assessment of qualitative parameters were good [0.74≤Kappa value or ICC≤0.92].
CONCLUSIONS: Rs-EPI DWI sequence is a potential technique to improve the imaging quality in the diagnosis of intracranial mass lesions.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app