A "quality-control-based correction method" for displayed dose indices on CT scanner consoles in patient dose surveys.

PURPOSE: A new quality-control-based (QC-based) method is introduced to obtain correction factors to be applied to displayed patient dose indices (CTDIVol and DLP) on CT scanner consoles to verify improvement of dose surveys for diagnostic reference levels (DRLs) determination.

METHOD: An available data-base of QC documents and reports of 57 CT scanners in Tehran, Iran was used to estimate CTDIVol , DLP and relevant correction factors for three CT examination types including head, chest and abdomen/pelvis. The correction factor is the ratio of QC-based estimated dose to displayed dose. A dose survey was performed by applying on-site "data collection method" and correction factors obtained in order to select CT scanners in three modes for determination of CT DRLs by inclusion of: (a) all CT scanners before displayed dose indices were corrected (57), (b) only CT scanners calibrated by QC experts (41) and (c) all CT scanners after displayed dose indices were corrected (57).

RESULTS: For the 41 CT scanners, correction factors of three examination types obtained in this study are within the acceptance tolerance of IAEA HHS-19. The correction factors range from 0.45 to 1.7 (average of 3 examinations) which is due to the change in the calibrated value of CTDIVol over extended time. The DRL differences in three modes are within ±1.0% for CTDIVol and ±12.4% for DLP.

CONCLUSIONS: The "QC-based correction method" applied to mode (c) has improved the DRLs obtained by other two modes. This method is a strong alternative to "direct dose measurement" with simplicity and cost effectiveness.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app