We have located links that may give you full text access.
Journal Article
Validation Studies
Generating ActiGraph Counts from Raw Acceleration Recorded by an Alternative Monitor.
Medicine and Science in Sports and Exercise 2017 November
PURPOSE: This study aimed to implement an aggregation method in Matlab for generating ActiGraph counts from raw acceleration recorded with an alternative accelerometer device and to investigate the validity of the method.
METHODS: The aggregation method, including the frequency band-pass filter, was implemented and optimized based on standardized sinusoidal acceleration signals generated in Matlab and processed in the ActiLife software. Evaluating the validity of the aggregation method was approached using a mechanical setup and with a 24-h free-living recording using a convenient sample of nine subjects. Counts generated with the aggregation method applied to Axivity AX3 raw acceleration data were compared with counts generated with ActiLife from ActiGraph GT3X+ data.
RESULTS: An optimal band-pass filter was fitted resulting in a root-mean-square error of 25.7 counts per 10 s and mean absolute error of 15.0 counts per second across the full frequency range. The mechanical evaluation of the proposed aggregation method resulted in an absolute mean ± SD difference of -0.11 ± 0.97 counts per 10 s across all rotational frequencies compared with the original ActiGraph method. Applying the aggregation method to the 24-h free-living recordings resulted in an epoch level bias ranging from -16.2 to 0.9 counts per 10 s, a relative difference in the averaged physical activity (counts per minute) ranging from -0.5% to 4.7% with a group mean ± SD of 2.2% ± 1.7%, and a Cohen's kappa of 0.945, indicating almost a perfect agreement in the intensity classification.
CONCLUSION: The proposed band-pass filter and aggregation method is highly valid for generating ActiGraph counts from raw acceleration data recorded with alternative devices. It would facilitate comparability between studies using different devices collecting raw acceleration data.
METHODS: The aggregation method, including the frequency band-pass filter, was implemented and optimized based on standardized sinusoidal acceleration signals generated in Matlab and processed in the ActiLife software. Evaluating the validity of the aggregation method was approached using a mechanical setup and with a 24-h free-living recording using a convenient sample of nine subjects. Counts generated with the aggregation method applied to Axivity AX3 raw acceleration data were compared with counts generated with ActiLife from ActiGraph GT3X+ data.
RESULTS: An optimal band-pass filter was fitted resulting in a root-mean-square error of 25.7 counts per 10 s and mean absolute error of 15.0 counts per second across the full frequency range. The mechanical evaluation of the proposed aggregation method resulted in an absolute mean ± SD difference of -0.11 ± 0.97 counts per 10 s across all rotational frequencies compared with the original ActiGraph method. Applying the aggregation method to the 24-h free-living recordings resulted in an epoch level bias ranging from -16.2 to 0.9 counts per 10 s, a relative difference in the averaged physical activity (counts per minute) ranging from -0.5% to 4.7% with a group mean ± SD of 2.2% ± 1.7%, and a Cohen's kappa of 0.945, indicating almost a perfect agreement in the intensity classification.
CONCLUSION: The proposed band-pass filter and aggregation method is highly valid for generating ActiGraph counts from raw acceleration data recorded with alternative devices. It would facilitate comparability between studies using different devices collecting raw acceleration data.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Diagnosis and Management of Cardiac Sarcoidosis: A Scientific Statement From the American Heart Association.Circulation 2024 April 19
Essential thrombocythaemia: A contemporary approach with new drugs on the horizon.British Journal of Haematology 2024 April 9
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app