Bolus Insulin Dose Error Distributions Based on Results From Two Clinical Trials Comparing Blood Glucose Monitoring Systems.

BACKGROUND: In 2 previous clinical trials, fingertip capillary blood samples were evaluated using prespecified blood glucose monitoring systems (BGMSs) and a reference YSI glucose analyzer. In post hoc analyses, hypothetical insulin doses were calculated using these blood glucose measurements; dosing errors were compared for each trial.

METHOD: For each blood glucose measurement, premeal bolus insulin dosing was determined for a hypothetical person, assuming a 60-g carbohydrate meal and 100-mg/dL target blood glucose level (adjusting 1/25 insulin sensitivity and 1/15 insulin:carbohydrate ratio inputs to account for BGMS measurement error). Dosing error was the difference between doses calculated using the BGMS and YSI results.

RESULTS: In Clinical Trial 1, 95% dose error ranges (in units of insulin) were: CONTOUR® NEXT EZ BGMS (EZ), -0.9 to 0.5; Accu-Chek® Aviva BGMS (ACA), -0.5 to 1.8; FreeStyle Freedom Lite® BGMS (FFL), -3.2 to -0.3; OneTouch® Ultra® 2 BGMS (OTU2), -4.1 to 0.3; and Truetrack® BGMS (TT), -3.9 to 2.2. In Clinical Trial 2, these ranges were: CONTOUR® NEXT BGMS (CN), -0.7 to 1.7; Accu-Chek® Aviva Nano BGMS (ACAN), -1.3 to 1.8; FreeStyle Lite® BGMS (FSL), -5.1 to 0.2; OTU2, -1.9 to 1.2; OneTouch® Verio® Pro BGMS (OTVP), -1.0 to 1.9; and TT, -5.1 to 1.7. Within each trial, EZ and CN had statistically significantly smaller insulin dose error ranges than other BGMSs ( P <0.0001).

CONCLUSIONS: The ranges of insulin dose errors were statistically significantly smaller with EZ and CN than with all other BGMSs in this post hoc analysis. Differences in BGMS accuracy could result in clinically important differences in insulin dosing.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app