Validation of a defibrillation lead ventricular volume measurement compared to three-dimensional echocardiography.
BACKGROUND: There is increasing evidence that using frequent invasive measures of pressure in patients with heart failure results in improved outcomes compared to traditional measures. Admittance, a measure of volume derived from preexisting defibrillation leads, is proposed as a new technique to monitor cardiac hemodynamics in patients with an implantable defibrillator.
OBJECTIVE: The purpose of this study was to evaluate the accuracy of a new ventricular volume sensor (VVS, CardioVol) compared with 3-dimenssional echocardiography (echo) in patients with an implantable defibrillator.
METHODS: Twenty-two patients referred for generator replacement had their defibrillation lead attached to VVS to determine the level of agreement to a volume measurement standard (echo). Two opposite hemodynamic challenges were sequentially applied to the heart (overdrive pacing and dobutamine administration) to determine whether real changes in hemodynamics could be reliably and repeatedly assessed with VVS. Equivalence of end-diastolic volume (EDV) and stroke volume (SV) determined by both methods was also assessed.
RESULTS: EDV and SV were compared using VVS and echo. VVS tracked expected physiologic trends. EDV was modulated -10% by overdrive pacing (14 mL). SV was modulated -13.7% during overdrive pacing (-6 mL) and increased over baseline +14.6% (+8 mL) with dobutamine. VVS and echo mean EDVs were found statistically equivalent, with margin of equivalence 13.8 mL (P <.05). Likewise, mean SVs were found statistically equivalent with margin of equivalence 15.8 mL (P <.05).
CONCLUSION: VVS provides an accurate method for ventricular volume assessment using chronically implanted defibrillator leads and is statistically equivalent to echo determination of mean EDV and SV.
OBJECTIVE: The purpose of this study was to evaluate the accuracy of a new ventricular volume sensor (VVS, CardioVol) compared with 3-dimenssional echocardiography (echo) in patients with an implantable defibrillator.
METHODS: Twenty-two patients referred for generator replacement had their defibrillation lead attached to VVS to determine the level of agreement to a volume measurement standard (echo). Two opposite hemodynamic challenges were sequentially applied to the heart (overdrive pacing and dobutamine administration) to determine whether real changes in hemodynamics could be reliably and repeatedly assessed with VVS. Equivalence of end-diastolic volume (EDV) and stroke volume (SV) determined by both methods was also assessed.
RESULTS: EDV and SV were compared using VVS and echo. VVS tracked expected physiologic trends. EDV was modulated -10% by overdrive pacing (14 mL). SV was modulated -13.7% during overdrive pacing (-6 mL) and increased over baseline +14.6% (+8 mL) with dobutamine. VVS and echo mean EDVs were found statistically equivalent, with margin of equivalence 13.8 mL (P <.05). Likewise, mean SVs were found statistically equivalent with margin of equivalence 15.8 mL (P <.05).
CONCLUSION: VVS provides an accurate method for ventricular volume assessment using chronically implanted defibrillator leads and is statistically equivalent to echo determination of mean EDV and SV.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app