Early Hemodynamic Results in Patients With Small Aortic Annulus After Aortic Valve Replacement.
OBJECTIVE: Patients with a small aortic annulus (≤21 mm) have an increased risk of patient-prosthesis mismatch after valve replacement. The aim of this study was to compare the early hemodynamic performance of the balloon-expandable transaortic valve implantation Edwards system (SAPIEN) and the sutureless Perceval prostheses.
METHODS: Fifty patients underwent transcatheter aortic valve implantation, and 113 patients underwent sutureless aortic valve replacement. Mean ± SD aortic annulus diameter was 19.7 ± 1 mm, with no significant difference between groups. SAPIEN valve size was 23 mm in 40 patients (80%) and 26 mm in 10 patients (20%). Perceval valve size was small in 45 patients (40%), medium in 62 patients (55%), and large in 6 patients (5%). Transthoracic Doppler echocardiographic images were collected at baseline and before discharge.
RESULTS: There were no significant difference in predischarge effective orifice area (SAPIEN: 1.5 ± 0.5 cm and Perceval: 1.48 ± 0.34 cm, P = 0.58) and indexed effective orifice areas (SAPIEN: 0.93 ± 0.32 cm/m and Perceval: 0.88 ± 0.22 cm/m, P = 0.42). Predischarge mean ± SD transaortic gradient was lower with the SAPIEN than with Perceval valves (12 ± 6 and 17 ± 6 mm Hg, respectively, P < 0.001). Rates of moderate and severe prosthesis-patient mismatch were similar (SAPIEN: 44% and 10% and Perceval: 50% and 14%, P = 0.53 and 0.75, respectively). There were no moderate-severe paravalvular leaks.
CONCLUSIONS: Although indexed effective orifice areas were similar, transcatheter aortic valve implantation with the balloon-expandable SAPIEN system yielded lower predischarge transaortic mean gradients than the surgically implanted Perceval, in patients with a small annulus.
METHODS: Fifty patients underwent transcatheter aortic valve implantation, and 113 patients underwent sutureless aortic valve replacement. Mean ± SD aortic annulus diameter was 19.7 ± 1 mm, with no significant difference between groups. SAPIEN valve size was 23 mm in 40 patients (80%) and 26 mm in 10 patients (20%). Perceval valve size was small in 45 patients (40%), medium in 62 patients (55%), and large in 6 patients (5%). Transthoracic Doppler echocardiographic images were collected at baseline and before discharge.
RESULTS: There were no significant difference in predischarge effective orifice area (SAPIEN: 1.5 ± 0.5 cm and Perceval: 1.48 ± 0.34 cm, P = 0.58) and indexed effective orifice areas (SAPIEN: 0.93 ± 0.32 cm/m and Perceval: 0.88 ± 0.22 cm/m, P = 0.42). Predischarge mean ± SD transaortic gradient was lower with the SAPIEN than with Perceval valves (12 ± 6 and 17 ± 6 mm Hg, respectively, P < 0.001). Rates of moderate and severe prosthesis-patient mismatch were similar (SAPIEN: 44% and 10% and Perceval: 50% and 14%, P = 0.53 and 0.75, respectively). There were no moderate-severe paravalvular leaks.
CONCLUSIONS: Although indexed effective orifice areas were similar, transcatheter aortic valve implantation with the balloon-expandable SAPIEN system yielded lower predischarge transaortic mean gradients than the surgically implanted Perceval, in patients with a small annulus.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app