Add like
Add dislike
Add to saved papers

CORRECTION FACTORS FOR ATTENUATION AND SCATTERING OF THE WALL OF A CYLINDRICAL IONIZATION CHAMBER IN THE 6-7 MeV HIGH-ENERGY PHOTON REFERENCE FIELD.

In high-energy photon reference fields the value of the air kerma rate is determined by using ionization chambers (ICs). From the charge collected inside the IC the dose can be calculated using a set of calibration and correction factors according to ISO 4037-2. A crucial parameter is the correction for the attenuation and scattering of the primary radiation due to the chamber wall. This parameter can be determined using Monte Carlo calculations. The evaluation of the factor was performed for a commercially available IC of the type Victoreen 550-3 under different build-up conditions. The results were verified by measurements in the R-F high-energy photon fields according to ISO 4037-1 at the Physikalisch-Technische Bundesanstalt (PTB) and the Japan Atomic Energy Agency (JAEA).

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app