A Prospective Clinical Trial Comparing Visible Light Spectroscopy to Handheld Doppler for Postoperative Free Tissue Transfer Monitoring.
Plastic and Reconstructive Surgery 2017 September
BACKGROUND: Early detection of compromised free flap perfusion is critical. A common modality of thrombosis detection, physical examination augmented with hand-held Doppler, provides only intermittent data and is insensitive to venous compromise. Visible light spectroscopy provides continuous, noninvasive evaluation of tissue perfusion. The authors hypothesized that visible light spectroscopy is a more sensitive and specific monitoring method for early detection of postoperative flap compromise than intermittent hand-held Doppler and clinical examination.
METHODS: The authors prospectively conducted a controlled study evaluating the sensitivity, specificity, and accuracy of the T-Stat Model 303 VLS oximeter versus that of intermittent Doppler and clinical examination. The authors prospectively collected and analyzed patient data, complications, reoperations, flap failures, and tissue oxygen saturation.
RESULTS: Sixty-eight patients with 81 flaps completed the study. The majority of flaps (86.4 percent) were either transverse rectus abdominis musculocutaneous or deep inferior epigastric artery perforator flaps. The mean tissue oxygen saturation for all flaps in the study was 56.7 ± 7.1 percent (range, 39.4 to 72.1 percent) and did not differ significantly with patient comorbidity or flap type. During the course of the study, three flaps were returned to the operating room for exploration because of a perfusion abnormality, and the resulting salvage rate was 100 percent. The sensitivity, specificity, and accuracy of visible light spectroscopy were found to be greater than both intermittent Doppler and clinical examination.
CONCLUSION: Visible light spectroscopy is a reliable, continuous adjunct to free tissue transfer monitoring with advantages over intermittent hand-held Doppler and clinical examination.
CLINICAL QUESTION/LEVEL OF EVIDENCE: Diagnostic, II.
METHODS: The authors prospectively conducted a controlled study evaluating the sensitivity, specificity, and accuracy of the T-Stat Model 303 VLS oximeter versus that of intermittent Doppler and clinical examination. The authors prospectively collected and analyzed patient data, complications, reoperations, flap failures, and tissue oxygen saturation.
RESULTS: Sixty-eight patients with 81 flaps completed the study. The majority of flaps (86.4 percent) were either transverse rectus abdominis musculocutaneous or deep inferior epigastric artery perforator flaps. The mean tissue oxygen saturation for all flaps in the study was 56.7 ± 7.1 percent (range, 39.4 to 72.1 percent) and did not differ significantly with patient comorbidity or flap type. During the course of the study, three flaps were returned to the operating room for exploration because of a perfusion abnormality, and the resulting salvage rate was 100 percent. The sensitivity, specificity, and accuracy of visible light spectroscopy were found to be greater than both intermittent Doppler and clinical examination.
CONCLUSION: Visible light spectroscopy is a reliable, continuous adjunct to free tissue transfer monitoring with advantages over intermittent hand-held Doppler and clinical examination.
CLINICAL QUESTION/LEVEL OF EVIDENCE: Diagnostic, II.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app