We have located links that may give you full text access.
Long-term results of total hip arthroplasty with the CementLess Spotorno (CLS) system.
Hip International : the Journal of Clinical and Experimental Research on Hip Pathology and Therapy 2017 September 20
BACKGROUND: This study presents the long-term results of the Cementless Spotorno (CLS) total hip arthroplasty system and an analysis of factors associated with clinical and radiographic outcome.
METHODS: We studied a series of 120 consecutive CLS arthroplasties in a young patient group (mean age at surgery: 55.9 ± 5.9 years). The Merle d'Aubigné-Postel score, polyethylene (PE) wear, and radiographic status were recorded during follow-up. Survival analyses, repeated-measures analysis of variance, and a nested case-control study were used for statistical evaluation.
RESULTS: After a mean follow-up of 14.6 years (range 0.1-24.2 years, including revisions and lost to follow-up), 24 revisions had been performed, 16 of which for aseptic cup loosening. Kaplan-Meier survival analysis showed a 24-year survival of 72.8% (95% CI, 63.0%-82.6%) with revision for any reason as endpoint, and 80.1% (95% CI, 70.9%-89.3%) for revision for aseptic cup loosening. Mean final Merle d'Aubigné-Postel score was 16.1 points (range 7-18). Mean PE wear at final follow-up was 2.3 mm (range 0.6-6.8 mm). A higher rate of PE wear was associated with better clinical scores but also with revision for cup loosening. Factors associated with more PE wear were: younger age at surgery; 32 - mm head; longer follow-up; and steeper inclination angle.
CONCLUSIONS: Beyond 10 years, the CLS stem is reliable, but the high revision rate for aseptic cup loosening is concerning, specifically with better performing (cementless) alternatives available.
METHODS: We studied a series of 120 consecutive CLS arthroplasties in a young patient group (mean age at surgery: 55.9 ± 5.9 years). The Merle d'Aubigné-Postel score, polyethylene (PE) wear, and radiographic status were recorded during follow-up. Survival analyses, repeated-measures analysis of variance, and a nested case-control study were used for statistical evaluation.
RESULTS: After a mean follow-up of 14.6 years (range 0.1-24.2 years, including revisions and lost to follow-up), 24 revisions had been performed, 16 of which for aseptic cup loosening. Kaplan-Meier survival analysis showed a 24-year survival of 72.8% (95% CI, 63.0%-82.6%) with revision for any reason as endpoint, and 80.1% (95% CI, 70.9%-89.3%) for revision for aseptic cup loosening. Mean final Merle d'Aubigné-Postel score was 16.1 points (range 7-18). Mean PE wear at final follow-up was 2.3 mm (range 0.6-6.8 mm). A higher rate of PE wear was associated with better clinical scores but also with revision for cup loosening. Factors associated with more PE wear were: younger age at surgery; 32 - mm head; longer follow-up; and steeper inclination angle.
CONCLUSIONS: Beyond 10 years, the CLS stem is reliable, but the high revision rate for aseptic cup loosening is concerning, specifically with better performing (cementless) alternatives available.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app