Enhanced quantitative susceptibility mapping (QSM) using real-time field control.
Magnetic Resonance in Medicine 2018 Februrary
PURPOSE: To assess the potential of a real-time field-control (FC) system for mitigating effects of spatiotemporal field fluctuations in quantitative susceptibility mapping (QSM) at 7 T.
METHODS: Magnitude, phase, and QSM images of phantoms and healthy volunteers were acquired under standard conditions and under induced field perturbation (FP) (phantoms: periodic water-bottle displacement; volunteers: deep breathing and forearm movement) with and without FC, which continuously detects and minimizes magnetic-field variations.
RESULTS: Field control successfully eliminated FP-induced impairment of phantom image quality and deviations from a linear susceptibility increase for increasing gadolinium concentration in a Gd dilution series (y = 320x - 0.60, R2 = 0.93 for the scan with FP and FC versus y = 259x - 0.54, R2 = 0.78 for the scan with FP and no FC (slope literature value: 326 ppm L/mol)). Similarly, in volunteers, FC allowed a recovery of a FP-induced loss of identifiable brain structures and reduced the relative change of mean susceptibilities and standard deviations (93 ± 53% to 34 ± 46%) in all regions of interests with respect to the reference scan.
CONCLUSIONS: Real-time FC improved the delineation of brain structures and the match of susceptibility values with reference values obtained without FP. Magn Reson Med 79:770-778, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
METHODS: Magnitude, phase, and QSM images of phantoms and healthy volunteers were acquired under standard conditions and under induced field perturbation (FP) (phantoms: periodic water-bottle displacement; volunteers: deep breathing and forearm movement) with and without FC, which continuously detects and minimizes magnetic-field variations.
RESULTS: Field control successfully eliminated FP-induced impairment of phantom image quality and deviations from a linear susceptibility increase for increasing gadolinium concentration in a Gd dilution series (y = 320x - 0.60, R2 = 0.93 for the scan with FP and FC versus y = 259x - 0.54, R2 = 0.78 for the scan with FP and no FC (slope literature value: 326 ppm L/mol)). Similarly, in volunteers, FC allowed a recovery of a FP-induced loss of identifiable brain structures and reduced the relative change of mean susceptibilities and standard deviations (93 ± 53% to 34 ± 46%) in all regions of interests with respect to the reference scan.
CONCLUSIONS: Real-time FC improved the delineation of brain structures and the match of susceptibility values with reference values obtained without FP. Magn Reson Med 79:770-778, 2018. © 2017 International Society for Magnetic Resonance in Medicine.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app