Add like
Add dislike
Add to saved papers

The FLIR ONE thermal imager for the assessment of burn wounds: Reliability and validity study.

Burns 2017 November
BACKGROUND: Objective measurement tools may be of great value to provide early and reliable burn wound assessment. Thermal imaging is an easy, accessible and objective technique, which measures skin temperature as an indicator of tissue perfusion. These thermal images might be helpful in the assessment of burn wounds. However, before implementation of a novel measurement tool into clinical practice is considered, it is appropriate to test its clinimetric properties (i.e. reliability and validity). The objective of this study was to assess the reliability and validity of the recently introduced FLIR ONE thermal imager.

MATERIAL AND METHODS: Two observers obtained thermal images of burn wounds in adult patients at day 1-3, 4-7 and 8-10 after burn. Subsequently, temperature differences between the burn wound and healthy skin (ΔT) were calculated on an iPad mini containing the FLIR Tools app. To assess reliability, ΔT values of both observers were compared by calculating the intraclass correlation coefficient (ICC) and measurement error parameters. To assess validity, the ΔT values of the first observer were compared to the registered healing time of the burn wounds, which was specified into three categories: (I) ≤14 days, (II) 15-21 days and (III) >21 days. The ability of the FLIR ONE to discriminate between healing ≤21 days and >21 days was evaluated by means of a receiver operating characteristic curve and an optimal ΔT cut-off value.

RESULTS: Reliability: ICCs were 0.99 for each time point, indicating excellent reliability up to 10 days after burn. The standard error of measurement varied between 0.17-0.22°C.

VALIDITY: the area under the curve was calculated at 0.69 (95% CI 0.54-0.84). A cut-off value of -1.15°C shows a moderate discrimination between burn wound healing ≤21 days and >21 days (46% sensitivity; 82% specificity).

CONCLUSION: Our results show that the FLIR ONE thermal imager is highly reliable, but the moderate validity calls for additional research. However, the FLIR ONE is pre-eminently feasible, allowing easy and fast measurements in clinical burn practice.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app