Does gantry rotation time influence accuracy of volume computed tomography dose index (CTDI vol ) in modern CT?
Physica Medica : PM 2017 May
PURPOSE: The purpose of this study was to develop a gantry overrun corrected CTDIvol (cCTDIvol ) dosimetry and evaluate the differences between the displayed CTDIvol (dCTDIvol ), measured CTDIvol (mCTDIvol ), and the cCTDIvol .
METHODS AND MATERIALS: The each 8 rotation times between 275 and 1000ms of two CT scanners were investigated. Rotation time (Trot ) and the beam-on time (Tbeam ) in axial scanning were measured accurately to determine the gantry overrun time (Tover ) as Tbeam -Trot . Subsequently, mCTDIvol was measured by using a 100mm ionization chamber and CTDI phantoms. Furthermore, we introduced a gantry overrun correction factor (Co =Trot /Tbeam ) to obtain cCTDIvol . Upon completion of the data acquisition, the dCTDIvol and mCTDIvol were compared with the cCTDIvol .
RESULTS: The discrepancies of Trot were 0.2±0.2ms as compared to the preset rotation times, and Tover was machine-specific and almost constant (22.4±0.5ms or 45.1±0.3ms) irrespective of the preset rotation time. Both dCTDIvol and mCTDIvol were increasingly overestimated compared to cCTDIvol as the faster the preset rotation time was selected (1.7-23.5%).
CONCLUSION: The rotation time influences the accuracy of CTDIvol in modern CT, and should be taken into consideration when assessing the radiation output in modern CT.
METHODS AND MATERIALS: The each 8 rotation times between 275 and 1000ms of two CT scanners were investigated. Rotation time (Trot ) and the beam-on time (Tbeam ) in axial scanning were measured accurately to determine the gantry overrun time (Tover ) as Tbeam -Trot . Subsequently, mCTDIvol was measured by using a 100mm ionization chamber and CTDI phantoms. Furthermore, we introduced a gantry overrun correction factor (Co =Trot /Tbeam ) to obtain cCTDIvol . Upon completion of the data acquisition, the dCTDIvol and mCTDIvol were compared with the cCTDIvol .
RESULTS: The discrepancies of Trot were 0.2±0.2ms as compared to the preset rotation times, and Tover was machine-specific and almost constant (22.4±0.5ms or 45.1±0.3ms) irrespective of the preset rotation time. Both dCTDIvol and mCTDIvol were increasingly overestimated compared to cCTDIvol as the faster the preset rotation time was selected (1.7-23.5%).
CONCLUSION: The rotation time influences the accuracy of CTDIvol in modern CT, and should be taken into consideration when assessing the radiation output in modern CT.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app