Dose warping performance in deformable image registration in lung.
Physica Medica : PM 2017 May
PURPOSE: It is unclear that spatial accuracy can reflect the impact of deformed dose distribution. In this study, we used dosimetric parameters to compare an in-house deformable image registration (DIR) system using NiftyReg, with two commercially available systems, MIM Maestro (MIM) and Velocity AI (Velocity).
METHODS: For 19 non-small-cell lung cancer patients, the peak inspiration (0%)-4DCT images were deformed to the peak expiration (50%)-4DCT images using each of the three DIR systems, which included computation of the deformation vector fields (DVF). The 0%-gross tumor volume (GTV) and the 0%-dose distribution were also then deformed using the DVFs. The agreement in the dose distributions for the GTVs was evaluated using generalized equivalent uniform dose (gEUD), mean dose (Dmean ), and three-dimensional (3D) gamma index (criteria: 3mm/3%). Additionally, a Dice similarity coefficient (DSC) was used to measure the similarity of the GTV volumes.
RESULTS: Dmean and gEUD demonstrated good agreement between the original and deformed dose distributions (differences were generally less than 3%) in 17 of the patients. In two other patients, the Velocity system resulted in differences in gEUD of 50.1% and 29.7% and in Dmean of 11.8% and 4.78%. The gamma index comparison showed statistically significant differences for the in-house DIR vs. MIM, and MIM vs. Velocity.
CONCLUSIONS: The finely tuned in-house DIR system could achieve similar spatial and dose accuracy to the commercial systems. Care must be taken, as we found errors of more than 5% for Dmean and 30% for gEUD, even with a commercially available DIR tool.
METHODS: For 19 non-small-cell lung cancer patients, the peak inspiration (0%)-4DCT images were deformed to the peak expiration (50%)-4DCT images using each of the three DIR systems, which included computation of the deformation vector fields (DVF). The 0%-gross tumor volume (GTV) and the 0%-dose distribution were also then deformed using the DVFs. The agreement in the dose distributions for the GTVs was evaluated using generalized equivalent uniform dose (gEUD), mean dose (Dmean ), and three-dimensional (3D) gamma index (criteria: 3mm/3%). Additionally, a Dice similarity coefficient (DSC) was used to measure the similarity of the GTV volumes.
RESULTS: Dmean and gEUD demonstrated good agreement between the original and deformed dose distributions (differences were generally less than 3%) in 17 of the patients. In two other patients, the Velocity system resulted in differences in gEUD of 50.1% and 29.7% and in Dmean of 11.8% and 4.78%. The gamma index comparison showed statistically significant differences for the in-house DIR vs. MIM, and MIM vs. Velocity.
CONCLUSIONS: The finely tuned in-house DIR system could achieve similar spatial and dose accuracy to the commercial systems. Care must be taken, as we found errors of more than 5% for Dmean and 30% for gEUD, even with a commercially available DIR tool.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app