Comparison of cycloplegic refraction between Grand Seiko autorefractor and Retinomax autorefractor in the Vision in Preschoolers-Hyperopia in Preschoolers (VIP-HIP) Study.
PURPOSE: To evaluate the agreement of cycloplegic refractive error measures between the Grand Seiko and Retinomax autorefractors in 4- and 5-year-old children.
METHODS: Cycloplegic refractive error of children was measured using the Grand Seiko and Retinomax during a comprehensive eye examination. Accommodative error was measured using the Grand Seiko. The differences in sphere, cylinder, spherical equivalent (SE) and intereye vector dioptric distance (VDD) between autorefractors were assessed using the Bland-Altman plot and 95% limits of agreement (95% LoA).
RESULTS: A total of 702 examinations were included. Compared to the Retinomax, the Grand Seiko provided statistically significantly larger values of sphere (mean difference, 0.34 D; 95% LoA, -0.46 to 1.14 D), SE (mean, 0.25 D; 95% LoA, -0.55 to 1.05 D), VDD (mean, 0.19 D; 95% LoA, -0.67 to 1.05 D), and more cylinder (mean, -0.18 D; 95% LoA, -0.91 to 0.55 D). The Grand Seiko measured ≥0.5 D than Retinomax in 43.1% of eyes for sphere and 29.8% of eyes for SE. In multivariate analysis, eyes with SE of >4 D (based on the average of two autorefractors) had larger differences in sphere (mean, 0.66 D vs 0.35 D; P < 0.0001) and SE (0.57 D vs 0.26 D; P < 0.0001) than eyes with SE of ≤4 D.
CONCLUSIONS: Under cycloplegia, the Grand Seiko provided higher measures of sphere, more cylinder, and higher SE than the Retinomax. Higher refractive error was associated with larger differences in sphere and SE between the Grand Seiko and Retinomax.
METHODS: Cycloplegic refractive error of children was measured using the Grand Seiko and Retinomax during a comprehensive eye examination. Accommodative error was measured using the Grand Seiko. The differences in sphere, cylinder, spherical equivalent (SE) and intereye vector dioptric distance (VDD) between autorefractors were assessed using the Bland-Altman plot and 95% limits of agreement (95% LoA).
RESULTS: A total of 702 examinations were included. Compared to the Retinomax, the Grand Seiko provided statistically significantly larger values of sphere (mean difference, 0.34 D; 95% LoA, -0.46 to 1.14 D), SE (mean, 0.25 D; 95% LoA, -0.55 to 1.05 D), VDD (mean, 0.19 D; 95% LoA, -0.67 to 1.05 D), and more cylinder (mean, -0.18 D; 95% LoA, -0.91 to 0.55 D). The Grand Seiko measured ≥0.5 D than Retinomax in 43.1% of eyes for sphere and 29.8% of eyes for SE. In multivariate analysis, eyes with SE of >4 D (based on the average of two autorefractors) had larger differences in sphere (mean, 0.66 D vs 0.35 D; P < 0.0001) and SE (0.57 D vs 0.26 D; P < 0.0001) than eyes with SE of ≤4 D.
CONCLUSIONS: Under cycloplegia, the Grand Seiko provided higher measures of sphere, more cylinder, and higher SE than the Retinomax. Higher refractive error was associated with larger differences in sphere and SE between the Grand Seiko and Retinomax.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app