Evaluation of the effect of temperature variation on response of PRESAGE® dosimeter.
Journal of Cancer Research and Therapeutics 2017 January
INTRODUCTION: Many factors, such as PRESAGE ® composition, dose rate, energy, and type of radiation, temperature, etc., may effect on PRESAGE ® dosimeter response. The aim of this study was investigating the effect of temperature variation on response of PRESAGE ® solid dosimeter.
MATERIALS AND METHODS: In this study, a PRESAGE ® solid detector was fabricated. Ninety-four percent weight polyurethane, 5% weight carbon tetrachloride, and 1% weight leucomalachite green were used. Radiological and physical characteristics of PRESAGEs ®, such as mass density, electron density, and effective number atomic were obtained and compared with water. Response of PRESAGE ® dosimeter in temperatures -4, 10, 25, 35, 45, 55, 65, 75, 85, and 90°C was evaluated. In addition, the absorption peak at various temperatures was investigated.
RESULTS: The results showed that the absorption peak at different temperatures was in the range of 630-635 nm. For temperatures below 75°C, the results indicated that temperature variation has no effect on the response of PRESAGE ® dosimeter whereas at the temperatures> 75°C, temperature variation has an effect on PRESAGE ® dosimeter response.
CONCLUSION: The finding showed that temperature changes have not impact on the absorption peak. In addition, the results related to the effect of temperature variation on the response of PRESAGE ® dosimeter showed that in the range of clinical applications (temperatures below 75°C), temperature variation has no effect on PRESAGE ® dosimeter response.
MATERIALS AND METHODS: In this study, a PRESAGE ® solid detector was fabricated. Ninety-four percent weight polyurethane, 5% weight carbon tetrachloride, and 1% weight leucomalachite green were used. Radiological and physical characteristics of PRESAGEs ®, such as mass density, electron density, and effective number atomic were obtained and compared with water. Response of PRESAGE ® dosimeter in temperatures -4, 10, 25, 35, 45, 55, 65, 75, 85, and 90°C was evaluated. In addition, the absorption peak at various temperatures was investigated.
RESULTS: The results showed that the absorption peak at different temperatures was in the range of 630-635 nm. For temperatures below 75°C, the results indicated that temperature variation has no effect on the response of PRESAGE ® dosimeter whereas at the temperatures> 75°C, temperature variation has an effect on PRESAGE ® dosimeter response.
CONCLUSION: The finding showed that temperature changes have not impact on the absorption peak. In addition, the results related to the effect of temperature variation on the response of PRESAGE ® dosimeter showed that in the range of clinical applications (temperatures below 75°C), temperature variation has no effect on PRESAGE ® dosimeter response.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app