Repeatability of topographic corneal thickness in keratoconus comparing Visante™ OCT and Oculus Pentacam HR ® topographer.
PURPOSE: To assess the repeatability and reproducibility of Visante™ OCT and Oculus Pentacam HR® in measuring central (CCT) and topographic corneal thickness (TCT) along the principle meridians in keratoconus (KC) corneae.
METHOD: Twenty participants diagnosed with KC were recruited. There were two study visits. On the first study visit, two repeated corneal thickness measurements were obtained with each instrument. Measurements were repeated at least 48hrs later in the same order. TCT were recorded in the 90, 180, 45 and 135 meridians at 1mm intervals across 8mm chord.
RESULTS: Mean CCT for Visante™ OCT was 484.97±43.14μm (range: 484.84-486.09) and Oculus Pentacam HR® was 478.86±45.31μm (range: 477.20-480.53). No significant difference in TCT between the two visits (p=0.54) and measurements (p=0.63) for Visante™ OCT. For Oculus Pentacam HR®, no significant difference was found for each visit (p=0.18) but differences existed in the measurements outside of the central region (p=0.001). Tukey post-hoc analysis shows the differences (p˂0.05) were found in the +1 and +4 (supero-temporal) locations in the 135 meridian. Significant differences were found comparing the two instruments, (p<0.05). Bland Altman plots were used to demonstrate the differences between the two instruments and indicate their limits of agreement.
CONCLUSION: Both instruments gave repeatable measurements as no significant differences were found in most locations in all meridians. Comparing the two instruments, they were not reproducible in all locations.
METHOD: Twenty participants diagnosed with KC were recruited. There were two study visits. On the first study visit, two repeated corneal thickness measurements were obtained with each instrument. Measurements were repeated at least 48hrs later in the same order. TCT were recorded in the 90, 180, 45 and 135 meridians at 1mm intervals across 8mm chord.
RESULTS: Mean CCT for Visante™ OCT was 484.97±43.14μm (range: 484.84-486.09) and Oculus Pentacam HR® was 478.86±45.31μm (range: 477.20-480.53). No significant difference in TCT between the two visits (p=0.54) and measurements (p=0.63) for Visante™ OCT. For Oculus Pentacam HR®, no significant difference was found for each visit (p=0.18) but differences existed in the measurements outside of the central region (p=0.001). Tukey post-hoc analysis shows the differences (p˂0.05) were found in the +1 and +4 (supero-temporal) locations in the 135 meridian. Significant differences were found comparing the two instruments, (p<0.05). Bland Altman plots were used to demonstrate the differences between the two instruments and indicate their limits of agreement.
CONCLUSION: Both instruments gave repeatable measurements as no significant differences were found in most locations in all meridians. Comparing the two instruments, they were not reproducible in all locations.
Full text links
Related Resources
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app