We have located links that may give you full text access.
Customized Hinged Covered Metallic Stents for the Treatment of Benign Main Bronchial Stenosis.
Annals of Thoracic Surgery 2017 August
BACKGROUND: To address the limitations of silicone stents, we designed a hinged self-expandable covered metallic stent. The aim of this study was to evaluate the safety and efficacy of the customized stents in clinical applications.
METHODS: This was a retrospective analysis. Under conscious sedation and local anesthesia, the stents were implanted or removed by interventional radiologists, with fluoroscopic guidance.
RESULTS: Of 24 patients with benign main bronchial stenosis, stents were successfully placed in 21 (87.5%). The low-pressure balloon before dilation failed in 1 case (4.17%) of left main bronchial cicatricial stenosis. In 2 other cases (8.33%), stent placement was abandoned. Stents were successfully removed between 29 and 103 days after the procedure. After stent removal, the follow-up lasted for at least 12 months. Restenosis occurred only in 1 case (4.55%) owing to bronchial collapse 3 days after stent removal. Dyspnea occurred in another case (4.55%) at 2 months after retrieval; recurrence was confirmed using bronchoscopy, leading to a left pneumonectomy.
CONCLUSIONS: The described procedure is safe and easy to be performed and avoids the use of intubation, bronchoscopy, and general anesthesia.
METHODS: This was a retrospective analysis. Under conscious sedation and local anesthesia, the stents were implanted or removed by interventional radiologists, with fluoroscopic guidance.
RESULTS: Of 24 patients with benign main bronchial stenosis, stents were successfully placed in 21 (87.5%). The low-pressure balloon before dilation failed in 1 case (4.17%) of left main bronchial cicatricial stenosis. In 2 other cases (8.33%), stent placement was abandoned. Stents were successfully removed between 29 and 103 days after the procedure. After stent removal, the follow-up lasted for at least 12 months. Restenosis occurred only in 1 case (4.55%) owing to bronchial collapse 3 days after stent removal. Dyspnea occurred in another case (4.55%) at 2 months after retrieval; recurrence was confirmed using bronchoscopy, leading to a left pneumonectomy.
CONCLUSIONS: The described procedure is safe and easy to be performed and avoids the use of intubation, bronchoscopy, and general anesthesia.
Full text links
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app