Effect of impact assembly on the interface deformation and fretting corrosion of modular hip tapers: An in vitro study.
Journal of Orthopaedic Research : Official Publication of the Orthopaedic Research Society 2018 January
Wear and corrosion at the modular head-neck junction has been recognised to be a potential clinical concern, with multiple reports on adverse local tissue reactions and subsequent early failure of metal-on-metal hip replacements. Furthermore, reports on head-neck taper corrosion are also being described with conventional metal-on-polyethylene bearings. Manufacturing tolerances, surgical technique, non-axial alignment, material combination, high frictional torque and high bending moment have all been implicated in the failure process. There is limited guidance on the force of impaction with which surgeons should assemble modular hip prostheses. This study aims to investigate the effect of impaction force on the deformation and corrosion of modular tapers. Short neck tapers with high surface roughness (average Rz = 16.58 μm, Ra = 4.14μm) and long neck tapers with low surface roughness (average Rz = 3.82 μm, Ra = 0.81μm), were assembled with CoCrMo alloy heads (smooth finish) under controlled conditions with 2, 4 or 8 kN of impaction force. Material combinations tested included CoCrMo-head/CoCrMo-neck and CoCrMo-head/Ti-6Al-4V-neck. Assessment of surface deformation before and after impaction was made using surface profilometry. Measurement of fretting current during sinusoidal cyclic loading evaluated mechanically assisted corrosion for each assembly load during short-term cyclic loading (1000-cycles) and long-term cyclic loading (5 million-cycles). Deformation on head and neck tapers increased with assembly load. Fretting currents during short term simulation testing showed significantly lower currents (p < 0.05), in 8 kN assemblies when compared to 2 and 4 kN, especially for the short-rough tapers. Long-term simulator testing demonstrated a progressive reduction in fretting corrosion for samples impacted with 4 and 8 kN; however, this reduction was greater for samples impacted at 8 kN even at the start of testing. Based on our results, surgeons could minimise mechanically assisted crevice corrosion by using higher impact loads when assembling the head to the stem in total hip arthroplasty. © 2017 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 36:405-416, 2018.
Full text links
Related Resources
Trending Papers
Heart failure with preserved ejection fraction: diagnosis, risk assessment, and treatment.Clinical Research in Cardiology : Official Journal of the German Cardiac Society 2024 April 12
Proximal versus distal diuretics in congestive heart failure.Nephrology, Dialysis, Transplantation 2024 Februrary 30
Efficacy and safety of pharmacotherapy in chronic insomnia: A review of clinical guidelines and case reports.Mental Health Clinician 2023 October
World Health Organization and International Consensus Classification of eosinophilic disorders: 2024 update on diagnosis, risk stratification, and management.American Journal of Hematology 2024 March 30
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app