We have located links that may give you full text access.
Determining fitting ranges of various bone conduction hearing aids.
Clinical Otolaryngology 2018 Februrary
OBJECTIVES: To define fitting ranges for nine bone conduction devices (BCDs) over different frequencies based on the device's maximum power output (MPO) and to validate the assessment of MPO of BCDs in the ear canal.
BACKGROUND: Maximum power output (MPO) is an important characteristic when fitting BCDs. It is the highest output level a device can deliver and is one of the major determinants of a device's fitting range. A skull simulator can be used to verify MPO of percutaneous BCDs. No such simulator is available for active and passive transcutaneous devices.
DESIGN: The MPO of nine different BCDs was assessed either by real-ear measurements and/or with skull simulator measurements.
MAIN OUTCOME MEASURES: MPO and cross-validation of the methods using the Bland-Altman method.
RESULTS: Percutaneous BCDs have higher MPO levels compared to active and passive transcutaneous devices. This results in a wide dynamic range of hearing for percutaneous devices. Moreover, the assessment of MPO by real-ear measurements was validated.
CONCLUSION: Based on MPO data, fitting ranges were defined for nine BCDs over seven frequencies.
BACKGROUND: Maximum power output (MPO) is an important characteristic when fitting BCDs. It is the highest output level a device can deliver and is one of the major determinants of a device's fitting range. A skull simulator can be used to verify MPO of percutaneous BCDs. No such simulator is available for active and passive transcutaneous devices.
DESIGN: The MPO of nine different BCDs was assessed either by real-ear measurements and/or with skull simulator measurements.
MAIN OUTCOME MEASURES: MPO and cross-validation of the methods using the Bland-Altman method.
RESULTS: Percutaneous BCDs have higher MPO levels compared to active and passive transcutaneous devices. This results in a wide dynamic range of hearing for percutaneous devices. Moreover, the assessment of MPO by real-ear measurements was validated.
CONCLUSION: Based on MPO data, fitting ranges were defined for nine BCDs over seven frequencies.
Full text links
Related Resources
Trending Papers
Challenges in Septic Shock: From New Hemodynamics to Blood Purification Therapies.Journal of Personalized Medicine 2024 Februrary 4
Molecular Targets of Novel Therapeutics for Diabetic Kidney Disease: A New Era of Nephroprotection.International Journal of Molecular Sciences 2024 April 4
The 'Ten Commandments' for the 2023 European Society of Cardiology guidelines for the management of endocarditis.European Heart Journal 2024 April 18
A Guide to the Use of Vasopressors and Inotropes for Patients in Shock.Journal of Intensive Care Medicine 2024 April 14
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app
All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.
By using this service, you agree to our terms of use and privacy policy.
Your Privacy Choices
You can now claim free CME credits for this literature searchClaim now
Get seemless 1-tap access through your institution/university
For the best experience, use the Read mobile app