Parallel-Stenting Technique in a Sandwich Configuration for Hypogastric Artery Preservation during Endovascular Aneurysm Repair: An In Vitro Study.

BACKGROUND: The aim of the study was to identify the best conditions in iliac sandwich procedure for hypogastric artery preservation during endovascular aneurysm repair, testing different devices, different oversizing (OS) degrees as well as different methods to measure it.

METHODS: Four external iliac devices (16-mm Endurant and 12-mm Aorfix limb extensions; 11- and 13-mm Viabahn endografts) were tested with 2 distinct internal iliac stent grafts (8-mm Advanta V12 and 8-mm Viabahn) inside different proximal silicon iliac limb models (10, 12, 14, 16, and 18 mm), simulating an iliac sandwich procedure for hypogastric preservation. After remodeling all devices in a saline bath at 37°C, the combinations were computed tomography scanned. Gutter size, parallel-stent compression, and inadequate parallel-stent deployment or infolding were recorded. Oversizing between both parallel stents and the iliac limb models were examined in terms of added diameter, perimeter, and area being additionally compared.

RESULTS: All three sizing methods (diameter, perimeter, and area) were highly correlated (diameter OS to perimeter and area OS correlation coefficient 0.998 and 0.997, respectively, P < 0.001 for both); thus, diameter OS was used for further comparisons. Increasing diameter OS (< 30%, 30-55%, 55-75%, and > 75%) showed a significant tendency toward smaller gutters (38.9, 12.2, 5.4, and 2.6 mm2 , respectively, P < 0.001) but also increasing parallel-stent compression (13.5%, 28.9%, 43.9%, and 55.1%, P < 0.001) and infolding (0%, 0%, 38%, and 60%, P < 0.001). There were no significant differences between the analyzed devices.

CONCLUSIONS: In iliac sandwich procedures, better apposition is usually achieved when using 30-55% diameter OS; higher OS is related to smaller gutters but higher rates of malpositioning and parallel-stent compression. No clear recommendations in material selection can be performed. All sizing methods are highly correlated and predictable.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app