Dosimetric Evaluation of a Flexible Dual Balloon-Constructed Applicator in Treating Anorectal Cancer.

BACKGROUND AND PURPOSE: To assess the dosimetric flexibility of a dual balloon brachytherapy applicator developed for the treatment of anorectal lesions.

MATERIALS AND METHODS: Different amounts of water were infused into the inner and outer balloon separately to study the asymmetrical distribution of the catheter, the radial distance of the active source channel to the inner surface of the global target volume , the space between the active source channels, and their dosimetric impact to target tissues and uninvolved rectum.

RESULTS: Increasing inner balloon volume directly increased both the space between the active source channels and the radial distance of the active source channel to the inner surface of the global target volume. The space between the active source channels and the percentage of global target volume received 150% or more of the prescribed dose to target had a strong inverse correlation (-0.881/ P = .007, -0.976/ P = .001, respectively) with the radial distance of the active source channel to the inner surface of the global target volume. Conformity index, dose to 2 cm3 of rectum, and total reference air kerma were strongly correlated with the radial distance of the active source channel to the inner surface of the global target volume, with values of 0.952 ( P = .001), 0.833 ( P = .015), and 0.922 ( P = .002), respectively. Percentage of global target volume received 150% or more of the prescribed dose was significantly correlated with the space between the active source channels (0.81/ P = .022), and conformity index was strongly inversely correlated with the space between the active source channels (-0.833/ P = .015).

CONCLUSION: The dual balloon-constructed Anorectal Applicator offers a flexible way to adjust the distances of the active source positions to the target in relation to uninvolved rectal wall. This flexibility simplifies planning which results in a highly conformal dose distribution to the target lesion while minimizing dose to normal rectal tissue.

Full text links

We have located links that may give you full text access.
Can't access the paper?
Try logging in through your university/institutional subscription. For a smoother one-click institutional access experience, please use our mobile app.

Related Resources

For the best experience, use the Read mobile app

Mobile app image

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app

All material on this website is protected by copyright, Copyright © 1994-2024 by WebMD LLC.
This website also contains material copyrighted by 3rd parties.

By using this service, you agree to our terms of use and privacy policy.

Your Privacy Choices Toggle icon

You can now claim free CME credits for this literature searchClaim now

Get seemless 1-tap access through your institution/university

For the best experience, use the Read mobile app